Replace the blood pump(s)<br />
10. The view “Operating mode” appears in the monitoring program. Select Univentricular or<br />
biventricular, then confirm the desired operating mode with . In biventricular<br />
mode, the view “Pump size and single-step mode” follows. In univentricular mode, the<br />
connector seal test follows first, ad then the view “Pump size and single-step mode”.<br />
11. Pull the seal plug out of the connector. When replacing the left pump: red marked connector;<br />
when replacing the right pump: blue marked connector! Insert the plug of the new driving<br />
tube into the connector that has become free. The plug snaps audibly into place.<br />
12. Check that the plug is firmly seated. To do so, pull the plug above the ribbed plug body.<br />
Never pull on the tube!<br />
13. Loosen the tube clamps on the cannulas.<br />
14. In the monitoring program, move the cursor to the single-step mode field.<br />
15. When replacing the left pump: Confirm “Step left” with to trigger a single-step.<br />
With , trigger additional individual steps until the left pump is completely deaired.<br />
If the pump is completely de-aired: Remove the de-airing needle. In the monitoring<br />
program, move the cursor to “Step right”, then confirm with .<br />
16. When replacing the right pump: Confirm “Step left” with to trigger a single-step.<br />
Move the cursor to “Step right”, then confirm with to trigger a single-step. With<br />
, trigger additional individual steps until the right pump is completely de-aired.<br />
If the pump is completely de-aired: Remove the de-airing needle.<br />
17. Move the cursor to OK, then confirm with . The system restarts with the set<br />
parameters.<br />
18. Observe the pump filling. If necessary, adjust the parameters.<br />
19. Secure all the connections with cable tie straps.<br />
12.1.3 Replacing the two blood pumps<br />
Material<br />
• 2 prepared replacement blood pumps (see Section 12.1.1, p. 76)<br />
• 1 tube attachment set from accessory set (cable tie strap, cable tie pliers)<br />
Warning!<br />
• If the left blood pump is replaced in biventricular mode, the right pump must also be<br />
stopped <strong>for</strong> the duration of the replacement to prevent overloading of the pulmonary<br />
circulation (risk of pulmonary edema).<br />
<strong>Instructions</strong><br />
1. Bring the patient into the Trendelenburg position.<br />
2. Carefully cut through the cable tie straps on all the inflow and outflow cannulas.<br />
Immediately check the integrity of all the cannulas.<br />
3. If necessary, log on to the monitoring program: Enter the <strong>use</strong>r number and password,<br />
then confirm with .<br />
4. Select the option “Drive pa<strong>use</strong>”, then confirm with . Confirm the decision in the<br />
dialog window with or . The <strong>Ikus</strong> stops.<br />
5. As soon as the pumps have stopped, disconnect the cannulas underneath the pumps<br />
and slide them off the pump.<br />
6. Check the cannulas <strong>for</strong> deposits. If necessary, remove the deposits carefully.<br />
7. Pull the driving tubes of the pumps out of the connectors. To do so, grab the plug by the<br />
ribbed plug body and pull it out of the connector.<br />
8. Close the connectors on the <strong>Ikus</strong> that have become free with seal plugs.<br />
9. Moisten the free cannula ends with sterile saline solution. Ensure that they are<br />
absolutely free of air. Connect the prepared replacement pumps to the cannulas.<br />
10. The view “Operating mode” appears in the monitoring program. Select biventricular,<br />
then confirm with . The “Pump size and single-step mode” view appears.<br />
11. Pull the seal plug out of both connectors on the <strong>Ikus</strong>.<br />
12. Insert the plugs of the new driving tubes into the connectors. While doing so, observe the<br />
colored markings: Insert the driving tube of the left pump into the red marked connector,<br />
and insert the driving tube of the right pump into the blue marked connector. The plugs<br />
snap audibly into place.<br />
13. Check that each plug is firmly seated. To do so, pull the plug above the ribbed plug body.<br />
1000002 Rev. 4.9.1 77