Intraoperative drive management<br />
Start up the <strong>Ikus</strong> and set the parameters<br />
6.2.2 Select the operating mode (“Operating mode” view)<br />
“Operating mode” view<br />
<strong>Instructions</strong><br />
1. In the operating room: Immediately restore the mains power supply.<br />
2. Move the cursor to univentricular or biventricular, then confirm the desired operating<br />
mode with .<br />
3. During biventricular mode: You arrive at the view “Pump size and single-step mode”<br />
(Section 6.2.3 on p. 34).<br />
During univentricular mode, the connector seal test follows: The <strong>Ikus</strong> checks that the blue<br />
marked driving tube connector is closed.<br />
Univentricular mode: Connector seal test<br />
<strong>Instructions</strong><br />
1. The message “Please close the right outlet” appears. Check the blue marked driving tube<br />
connector. If the connector is open: Close it with the seal plug.<br />
2. Move the cursor to OK, then confirm the message with . The <strong>Ikus</strong> checks that<br />
the connector is actually closed. If this is the case, you arrive at the view “Pump size and<br />
single-step mode”. Otherwise, the <strong>Ikus</strong> repeats the connector seal test.<br />
Note<br />
• The <strong>Ikus</strong> repeats the connector seal test up to three times. Then the system switches<br />
itself off. Then contact <strong>Berlin</strong> <strong>Heart</strong> GmbH service.<br />
6.2.3 Selecting the pump size<br />
<strong>Instructions</strong><br />
1. In the popup menu, select the desired pump size with ↓,↑ .<br />
2. During biventricular mode: Move the cursor with ← / → to the field “right pump”, then<br />
select the desired pump size with ↓,↑ .<br />
Popup menu <strong>for</strong> pump size selection<br />
Note<br />
• The selected size and rate of the right blood pump cannot be larger than the left pump.<br />
The list of the pump sizes <strong>for</strong> the right pump is restricted accordingly.<br />
34 1000002 Rev. 4.9.1