INSTRUCTION<br />

1. If indicated, start fibrillation.<br />

2. Excision at the LV apex. The ideal<br />

site is the lateral wall of the left ventricle<br />

near the apex. Distance<br />

between LAD (septum) and center of<br />

the incision: approx. 2 cm in children,<br />

approx. 3 to 4 cm in adults.<br />

See Fig. 15, page 17. Make a cylindrical<br />

excision. Select the diameter<br />

of the excision so that it is slightly<br />

smaller than the cannula diameter.<br />

Start the incision remote from the<br />

septum.<br />

3. Inspect the incision edges and the<br />

area surrounding the incision <strong>for</strong> any<br />

thrombi and residual trabeculae and<br />

excise these completely.<br />

Fig. 15 Optimum cannula position<br />

Fig. 14 Anastomosis of inflow cannula<br />

with LV apex<br />

1 approx. 2 cm (children)<br />

approx. 4 cm (adults)<br />

2 Start of the incision remote from the septum<br />

1000095x14 Rev. 8 17/28