Electromagnetic immunity - Part 1<br />
EMC tables<br />
14.2 Electromagnetic immunity - Part 1<br />
Noise immunity<br />
tests<br />
Discharges of static<br />
electricity (ESD)<br />
according to<br />
IEC 61000-4-2<br />
Fast transient<br />
electrical<br />
disturbance<br />
variables/bursts<br />
according to<br />
IEC 61000-4-4<br />
Surge voltages<br />
according to<br />
IEC 61000-4-5<br />
Voltage dips, shortterm<br />
interruptions<br />
and fluctuations of<br />
the power supply<br />
according to<br />
IEC 61000-4-11<br />
Magnetic field at the<br />
supply frequency<br />
(50/60 Hz)<br />
according to<br />
IEC 61000-4-8<br />
The EXCOR stationary Driving Unit <strong>Ikus</strong> system is intended <strong>for</strong> operation in the electromagnetic<br />
environment described below. The customer or <strong>use</strong>r of the EXCOR stationary<br />
Driving Unit <strong>Ikus</strong> system must ensure that it is <strong>use</strong>d in such an environment.<br />
IEC 60601 testing level Compliance level<br />
± 6 kV contact discharge<br />
± 8 kV air discharge<br />
± 2 kV <strong>for</strong> mains<br />
power lines<br />
± 1 kV <strong>for</strong> input and<br />
output lines<br />
± 1 kV alternate<br />
mode voltage<br />
± 2 kV synchronous<br />
mode voltage<br />
< 5 % UT (>95 % dip in the UT) <strong>for</strong> ½ period<br />
40 % U T<br />
60 % dip in the U T )<br />
<strong>for</strong> 5 periods<br />
70 % U T<br />
30 % dip in the U T)<br />
<strong>for</strong> 25 periods<br />
< 5 % U T<br />
(>95 % dip in the U T )<br />
<strong>for</strong> 5 seconds<br />
± 6 kV contact discharge<br />
± 8 kV air discharge<br />
± 2 kV <strong>for</strong> mains<br />
power lines<br />
± 1 kV <strong>for</strong> input and<br />
output lines<br />
± 1 kV alternate<br />
mode voltage<br />
± 2 kV synchronous<br />
mode voltage<br />
< 5 % UT (>95 % dip in the UT) <strong>for</strong> ½ period<br />
40 % U T<br />
60 % dip in the U T )<br />
<strong>for</strong> 5 periods<br />
70 % U T<br />
30 % dip in the U T)<br />
<strong>for</strong> 25 periods<br />
< 5 % U T<br />
(>95 % dip in the U T )<br />
<strong>for</strong> 5 seconds<br />
Electromagnetic environment -<br />
guidelines<br />
Floors should be made from wood or<br />
concrete or be equipped with ceramic<br />
tiles. If the floor is equipped with a<br />
synthetic material, the relative<br />
atmospheric humidity must be<br />
at least 30%.<br />
The quality of the supply voltage should<br />
correspond to a typical commercial and<br />
hospital environment.<br />
The quality of the supply voltage should<br />
correspond to a typical commercial and<br />
hospital environment.<br />
The quality of the supply voltage should<br />
correspond to a typical commercial and<br />
hospital environment. If the <strong>use</strong>r of the<br />
EXCOR stationary Driving Unit <strong>Ikus</strong><br />
system requires continued functionality<br />
even when interruptions in the energy<br />
supply occur, it is recommended to<br />
supply the EXCOR stationary Driving<br />
Unit <strong>Ikus</strong> system from an<br />
uninterruptible power supply or a<br />
battery.<br />
3 A/m 3 A/m Magnetic fields at the mains power<br />
frequency should correspond to<br />
the typical values encountered in<br />
commercial and hospital environments.<br />
NOTE U T is the mains power AC voltage be<strong>for</strong>e the application of the testing level.<br />