Electromagnetic immunity - Part 1<br />

EMC tables<br />

14.2 Electromagnetic immunity - Part 1<br />

Noise immunity<br />

tests<br />

Discharges of static<br />

electricity (ESD)<br />

according to<br />

IEC 61000-4-2<br />

Fast transient<br />

electrical<br />

disturbance<br />

variables/bursts<br />

according to<br />

IEC 61000-4-4<br />

Surge voltages<br />

according to<br />

IEC 61000-4-5<br />

Voltage dips, shortterm<br />

interruptions<br />

and fluctuations of<br />

the power supply<br />

according to<br />

IEC 61000-4-11<br />

Magnetic field at the<br />

supply frequency<br />

(50/60 Hz)<br />

according to<br />

IEC 61000-4-8<br />

The EXCOR stationary Driving Unit <strong>Ikus</strong> system is intended <strong>for</strong> operation in the electromagnetic<br />

environment described below. The customer or <strong>use</strong>r of the EXCOR stationary<br />

Driving Unit <strong>Ikus</strong> system must ensure that it is <strong>use</strong>d in such an environment.<br />

IEC 60601 testing level Compliance level<br />

± 6 kV contact discharge<br />

± 8 kV air discharge<br />

± 2 kV <strong>for</strong> mains<br />

power lines<br />

± 1 kV <strong>for</strong> input and<br />

output lines<br />

± 1 kV alternate<br />

mode voltage<br />

± 2 kV synchronous<br />

mode voltage<br />

< 5 % UT (>95 % dip in the UT) <strong>for</strong> ½ period<br />

40 % U T<br />

60 % dip in the U T )<br />

<strong>for</strong> 5 periods<br />

70 % U T<br />

30 % dip in the U T)<br />

<strong>for</strong> 25 periods<br />

< 5 % U T<br />

(>95 % dip in the U T )<br />

<strong>for</strong> 5 seconds<br />

± 6 kV contact discharge<br />

± 8 kV air discharge<br />

± 2 kV <strong>for</strong> mains<br />

power lines<br />

± 1 kV <strong>for</strong> input and<br />

output lines<br />

± 1 kV alternate<br />

mode voltage<br />

± 2 kV synchronous<br />

mode voltage<br />

< 5 % UT (>95 % dip in the UT) <strong>for</strong> ½ period<br />

40 % U T<br />

60 % dip in the U T )<br />

<strong>for</strong> 5 periods<br />

70 % U T<br />

30 % dip in the U T)<br />

<strong>for</strong> 25 periods<br />

< 5 % U T<br />

(>95 % dip in the U T )<br />

<strong>for</strong> 5 seconds<br />

Electromagnetic environment -<br />

guidelines<br />

Floors should be made from wood or<br />

concrete or be equipped with ceramic<br />

tiles. If the floor is equipped with a<br />

synthetic material, the relative<br />

atmospheric humidity must be<br />

at least 30%.<br />

The quality of the supply voltage should<br />

correspond to a typical commercial and<br />

hospital environment.<br />

The quality of the supply voltage should<br />

correspond to a typical commercial and<br />

hospital environment.<br />

The quality of the supply voltage should<br />

correspond to a typical commercial and<br />

hospital environment. If the <strong>use</strong>r of the<br />

EXCOR stationary Driving Unit <strong>Ikus</strong><br />

system requires continued functionality<br />

even when interruptions in the energy<br />

supply occur, it is recommended to<br />

supply the EXCOR stationary Driving<br />

Unit <strong>Ikus</strong> system from an<br />

uninterruptible power supply or a<br />

battery.<br />

3 A/m 3 A/m Magnetic fields at the mains power<br />

frequency should correspond to<br />

the typical values encountered in<br />

commercial and hospital environments.<br />

NOTE U T is the mains power AC voltage be<strong>for</strong>e the application of the testing level.<br />

102 1000002 Rev. 4.9.1