WARNING<br />

When using a connecting set: observe the specified minimum lengths<br />

of the connecting tube. See Tab. 1, page 16. If necessary: shorten the<br />

cannulae to comply with the minimum length of the connecting tube.<br />

Article No. Change in diameter Minimum length after<br />

cutting<br />

A12-016 16 to 12 mm 90 mm<br />

A09-012 12 to 9 mm 75 mm<br />

A06-009 9 to 6 mm 60 mm<br />

Tab. 1 Connecting set: minimum lengths of the connecting tube<br />

Anastomosis of inflow cannula with apex<br />

8.2.1 Anastomosis of inflow cannula with apex<br />

Changes:<br />

• The section has been supplemented.<br />

Make sure that the long side of the cannula tip lies remote from the<br />

septum and parallel to the lateral wall. See Fig. 15, page 17. Otherwise<br />

there is a risk of myocardial suction of the septal wall.<br />

IMPORTANT: the flow direction arrow on the cannula body is aligned<br />

with the long side of the cannula tip (except <strong>for</strong> apex cannulae C10A-<br />

030, C14A-040 and C18A-020).<br />

16/28 1000095x14 Rev. 8