WARNING<br />
When using a connecting set: observe the specified minimum lengths<br />
of the connecting tube. See Tab. 1, page 16. If necessary: shorten the<br />
cannulae to comply with the minimum length of the connecting tube.<br />
Article No. Change in diameter Minimum length after<br />
cutting<br />
A12-016 16 to 12 mm 90 mm<br />
A09-012 12 to 9 mm 75 mm<br />
A06-009 9 to 6 mm 60 mm<br />
Tab. 1 Connecting set: minimum lengths of the connecting tube<br />
Anastomosis of inflow cannula with apex<br />
8.2.1 Anastomosis of inflow cannula with apex<br />
Changes:<br />
• The section has been supplemented.<br />
Make sure that the long side of the cannula tip lies remote from the<br />
septum and parallel to the lateral wall. See Fig. 15, page 17. Otherwise<br />
there is a risk of myocardial suction of the septal wall.<br />
IMPORTANT: the flow direction arrow on the cannula body is aligned<br />
with the long side of the cannula tip (except <strong>for</strong> apex cannulae C10A-<br />
030, C14A-040 and C18A-020).<br />
16/28 1000095x14 Rev. 8