Battery f<strong>use</strong><br />

Detecting and eliminating problems<br />

Battery f<strong>use</strong> triggered in battery mode<br />

Warning!<br />

• The <strong>Ikus</strong> stops immediately. The blood pumps are no longer driven.<br />

• As long as the <strong>Ikus</strong> is in battery mode, a second person trained in the <strong>use</strong> of the manual<br />

pump must be constantly present with the patient.<br />

<strong>Instructions</strong><br />

1. Immediately supply the patient with the manual pump.<br />

Notify hotline! +49(0)30 8187 2772<br />

Note<br />

• If the battery f<strong>use</strong> is triggered in battery mode, the <strong>Ikus</strong> generates an audible alarm.<br />

• If the power supply of the <strong>Ikus</strong> is restored within five minutes by pressing the battery f<strong>use</strong><br />

button, the <strong>Ikus</strong> continues working with the last set parameters. If the <strong>Ikus</strong> is only<br />

restarted after more than five minutes, the <strong>Ikus</strong> runs through the start test and all of its<br />

parameters must be set again (duration approx. 10 minutes).<br />

84 1000002 Rev. 4.9.1