Battery f<strong>use</strong><br />
Detecting and eliminating problems<br />
Battery f<strong>use</strong> triggered in battery mode<br />
Warning!<br />
• The <strong>Ikus</strong> stops immediately. The blood pumps are no longer driven.<br />
• As long as the <strong>Ikus</strong> is in battery mode, a second person trained in the <strong>use</strong> of the manual<br />
pump must be constantly present with the patient.<br />
<strong>Instructions</strong><br />
1. Immediately supply the patient with the manual pump.<br />
Notify hotline! +49(0)30 8187 2772<br />
Note<br />
• If the battery f<strong>use</strong> is triggered in battery mode, the <strong>Ikus</strong> generates an audible alarm.<br />
• If the power supply of the <strong>Ikus</strong> is restored within five minutes by pressing the battery f<strong>use</strong><br />
button, the <strong>Ikus</strong> continues working with the last set parameters. If the <strong>Ikus</strong> is only<br />
restarted after more than five minutes, the <strong>Ikus</strong> runs through the start test and all of its<br />
parameters must be set again (duration approx. 10 minutes).<br />