25.10.2012 Views

Instructions for use: Ikus - Berlin Heart

Instructions for use: Ikus - Berlin Heart

Instructions for use: Ikus - Berlin Heart

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

12 Detecting and eliminating problems<br />

Possible problems<br />

Problem Ca<strong>use</strong> / Corrective action<br />

Visible defects on the <strong>Ikus</strong> In<strong>for</strong>m 0700-berlinheart<br />

Visible defects on the pump Replace the pump: see page 75<br />

For all questions / uncertainties:<br />

Notify hotline! +49(0)30 8187 2772<br />

Replace the blood pump(s)<br />

Deposits in the pump If deposits are beginning: Check the anticoagulation therapy and adjust it if necessary<br />

In the case of floating deposits: Replace the pump: see page 75<br />

The pump fills/empties<br />

incorrectly<br />

<strong>Ikus</strong>: Curves are at a<br />

standstill, parameters<br />

cannot be adjusted<br />

Curves are still at a<br />

standstill<br />

Audible and visible alarm<br />

from the <strong>Ikus</strong>, message<br />

“Error: no data from the<br />

control processor” or<br />

“Error: no reaction from the<br />

control processor”<br />

Note<br />

• If you must contact <strong>Berlin</strong> <strong>Heart</strong> service due to a malfunction: If possible, <strong>use</strong> the<br />

“EXCOR VAD emergency log” (standard <strong>for</strong>m see page 96). This log records relevant<br />

in<strong>for</strong>mation that can contribute to identifying the malfunction.<br />

12.1 Replace the blood pump(s)<br />

Necessary if<br />

• a visible defect exists in the pump(s)<br />

• floating deposits are found in the pump(s)<br />

Recommendation<br />

Evaluate the status of the patient and the hemodynamics<br />

Adjust the parameters if necessary<br />

Consider changing the cannula position; see page 66<br />

Possible ca<strong>use</strong>s<br />

- Communication error between control processor and laptop<br />

- Batteries supply insufficient current<br />

- Electronics have failed (main and backup control processors)<br />

What should be done?<br />

Switch the laptop off and on, wait during the start process, then restart the<br />

monitoring program. Important: The <strong>Ikus</strong> continues operating during this time with<br />

the set parameters!<br />

The <strong>Ikus</strong> works in emergency pulse mode (synchronous mode, systolic pressure<br />

200 mmHg, diastolic pressure -40 mmHg, rate 90 bpm, relative systolic duration 40%).<br />

In<strong>for</strong>m 0700-berlinheart. Restart the <strong>Ikus</strong> after discussing with service: see page 79<br />

Possible ca<strong>use</strong>s:<br />

- Simultaneous failure of both control processors<br />

- Power supply failure<br />

What should be done?<br />

Evaluate the status of the patient and the hemodynamics<br />

In<strong>for</strong>m 0700-berlinheart immediately.<br />

We recommend also replacing the driving tubes each time the pump is replaced (in<br />

univentricular mode red or blue) (see section 13.2, “Technical specifications”, p. 89 and<br />

section 11.5.1, “Replace the driving tube”, p. 65), unless the last replacement was very<br />

recent.<br />

1000002 Rev. 4.9.1 75

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!