WARNING<br />
The EXCOR system and its components may be <strong>use</strong>d only by<br />
prescription of the attending physician.<br />
Only system components may be <strong>use</strong>d with the EXCOR system. No<br />
components other than those supplied by <strong>Berlin</strong> <strong>Heart</strong> may be <strong>use</strong>d.<br />
The warranty shall otherwise be voided.<br />
The long-term storage conditions <strong>for</strong> all sterile products must be<br />
observed: temperature +15°C to 25°C, relative humidity 35% to 50%.<br />
Store in a dry place! Otherwise there is a risk that the products may no<br />
longer be sterile.<br />
Do not resterilize or re<strong>use</strong> EXCOR sterile components! Otherwise<br />
there is a risk of infection or damage to materials. EXCOR sterile components<br />
are: blood pumps, cannulae, drivelines, accessory set (de-airing<br />
set, tube connecting set, membrane set, pressure gauge set).<br />
Do not resterilize or re<strong>use</strong> EXCOR sterile components after the <strong>use</strong>-by<br />
date. Otherwise there is a risk of infection or damage to materials. The<br />
<strong>use</strong>-by date is printed on the labels of the inner and outer packaging.<br />
Only remove the de-airing needle if the blood pump is running and the<br />
parameters have been adjusted.<br />
During implantation: ensure that the blood pumps are completely free<br />
of air at the start. Do this by first triggering a single step, then if<br />
necessary expelling any air bubbles using the de-airing needle. See<br />
section 6.2.6 of the instructions <strong>for</strong> <strong>use</strong>.<br />
Do not reintroduce de-airing needles that have been once removed.<br />
Otherwise the membrane may be damaged. See section 6.2.5 of the<br />
instructions <strong>for</strong> <strong>use</strong>.<br />
Do not touch the pumps or cannulae under any circumstances with<br />
pointed or sharp objects (e.g. surgical instruments). See chapter 8 of<br />
the instructions <strong>for</strong> <strong>use</strong>.<br />