During implantation: ensure that the connection between the pump<br />
and cannula is stable and airtight. Always secure each connection with<br />
at least one cable tie, using a tie-band gun. See section 8.8 of the<br />
instructions <strong>for</strong> <strong>use</strong>.<br />
Immediately after implantation: visually inspect the cannulae and their<br />
connections.<br />
Post-operative care: check the blood pumps and the visible parts of<br />
the cannulae at least 3 times a day <strong>for</strong> thrombi and deposits. Check<br />
the cannulae at least once a day <strong>for</strong> damage. See section 9.5 of the<br />
instructions <strong>for</strong> <strong>use</strong>.<br />
Post-operative care: ensure that the pumps, cannulae and drivelines<br />
are never exposed to tension, version and kinks. Otherwise there is a<br />
risk that the patient will receive inadequate support. Use a mirror to<br />
avoid unnecessary kinking in the cannulae when visually inspecting<br />
the underside of the pump.<br />
Cleaning: do not <strong>use</strong> acetone or mineral oil products in the immediate<br />
vicinity of the blood pump and driveline. Do not <strong>use</strong> corrosive agents,<br />
solutions containing dyes, or organic solvents <strong>for</strong> cleaning the blood<br />
pump and drivelines, since they may alter the surface of the product.<br />
We recommend using only water or alcohol to clean the blood pump<br />
and driveline.<br />
Cleaning: Do not <strong>use</strong> corrosive agents, solutions containing dyes, or<br />
organic solvents <strong>for</strong> cleaning the cannulae and transcutaneous exit<br />
site, since they may alter the surface of the product.<br />
We recommend using chlorhexidine solution to clean the cannulae<br />
and transcutaneous exit site.<br />
Monitoring program: To save data, <strong>use</strong> only the data storage media<br />
supplied (disks/USB sticks). Otherwise there is a risk that the data<br />
storage medium is not recognized and data are lost.<br />
End the monitor program only in an emergency. Since the text of a<br />
message is displayed only in the monitor program, there is no possibility<br />
of determining the type of the message when the monitor program<br />
is ended. When the monitor program is shut down, the only indications<br />
that there is an error message are an acoustic signal and the fact that<br />
the indicator lamp on the <strong>Ikus</strong> handle lights up.<br />