Replacing the driveline<br />
11.5.1 Replacing the driveline<br />
Changes:<br />
• The section has been supplemented.<br />
WARNING<br />
Check drivelines regularly by visual inspection <strong>for</strong> irregularities or<br />
damage. Replace drivelines immediately if damaged.<br />
The drivelines (red/blue) have a maximum period of <strong>use</strong> of 1 year. After this period,<br />
they must be replaced.<br />
Messages<br />
11.8 Please check left/right pump and tube<br />
Changes:<br />
• The section has been supplemented.<br />
IMPORTANT: if this message appears during biventricular operation in one pump and<br />
the <strong>use</strong>r stops the other pump, the message will be deleted during the subsequent<br />
start test, irrespective of whether the ca<strong>use</strong> of the message still exists.<br />
11.15 The error is only detected by the active (or passive) computer<br />
Changes:<br />
• The section has been supplemented.<br />
Only the active computer has detected an error. This message only appears in conjunction<br />
with an additional, actual error message. There may be a delay of several<br />
seconds between the two messages. It is essential to wait <strong>for</strong> the second message to<br />
appear. Immediately take all of the measures required by the second message.<br />
11.20 Processor failure. Contact customer service!<br />
Changes:<br />
• The section has been supplemented.<br />
IMPORTANT: it is possible that this message may not be displayed depending on the<br />
time of occurrence of the error. <strong>Ikus</strong> will then raise an alarm only visually and acoustically.<br />