14 EMC tables<br />

14.1 Emissions characteristics<br />

Measurement of transient<br />

emissions<br />

HF emissions according to<br />

CISPR 11<br />

HF emissions according to<br />

CISPR 11<br />

Emissions of harmonics<br />

according to IEC 61000-3-2<br />

Emissions of voltage<br />

fluctuations/flickers according<br />

to IEC 61000-3-3<br />

Emissions characteristics<br />

The EXCOR stationary Driving Unit <strong>Ikus</strong> system is intended <strong>for</strong> operation in the environment<br />

described below. The customer or <strong>use</strong>r of the EXCOR stationary Driving Unit <strong>Ikus</strong><br />

system must ensure that it is <strong>use</strong>d in such an environment.<br />

Compliance Electromagnetic environment - guidelines<br />

Group 1 The EXCOR stationary Driving Unit <strong>Ikus</strong> system <strong>use</strong>s<br />

HF energy only <strong>for</strong> its internal functioning. There<strong>for</strong>e,<br />

its HF emissions are very low, and it is improbable that<br />

interference with adjacent electronic devices will occur.<br />

Class B<br />

Class A<br />

Con<strong>for</strong>mant<br />

The EXCOR stationary Driving Unit <strong>Ikus</strong> system is suitable<br />

<strong>for</strong> <strong>use</strong> in all facilities including residential areas, etc.,<br />

which are directly connected to a public power supply<br />

which also supplies buildings <strong>use</strong>d <strong>for</strong> residential<br />

purposes.<br />

1000002 Rev. 4.9.1 101