Start menu<br />

5.1 Start menu<br />

Changes:<br />

• The section has been supplemented.<br />

4. Save data The LOG files can be saved on a data storage medium<br />

(disk/USB stick) with this option. This should be done<br />

only after consulting the Service department.<br />

Unintended change of the operating mode<br />

5.3.1 Pa<strong>use</strong> drive: stop the <strong>Ikus</strong> temporarily<br />

Changes:<br />

• The section has been supplemented.<br />

When restarting the system from the L/R separate mode there can be an unintentional<br />

switch to the synchronous mode, although L/R separate continues to be displayed.<br />

The pressure curves, however, will distinctly indicate the synchronous mode.<br />

In this case:<br />

INSTRUCTION<br />

1. Navigate the cursor to Synchronous mode. Do not confirm with .<br />

2. Navigate cursor to L/R separate. Do not confirm with .<br />

3. Navigate cursor to an arbitrary pump parameter. Confirm with .<br />

5.5 Switching from univentricular to biventricular operation<br />

Changes:<br />

• The section has been supplemented.<br />

In the change from univentricular to biventricular support, an unintentional change<br />

into the synchronous or separate mode can occur.<br />

In this case:<br />

INSTRUCTION<br />

1. Select the desired mode anew.<br />

2. Check the parameters.<br />