Start menu<br />
5.1 Start menu<br />
Changes:<br />
• The section has been supplemented.<br />
4. Save data The LOG files can be saved on a data storage medium<br />
(disk/USB stick) with this option. This should be done<br />
only after consulting the Service department.<br />
Unintended change of the operating mode<br />
5.3.1 Pa<strong>use</strong> drive: stop the <strong>Ikus</strong> temporarily<br />
Changes:<br />
• The section has been supplemented.<br />
When restarting the system from the L/R separate mode there can be an unintentional<br />
switch to the synchronous mode, although L/R separate continues to be displayed.<br />
The pressure curves, however, will distinctly indicate the synchronous mode.<br />
In this case:<br />
INSTRUCTION<br />
1. Navigate the cursor to Synchronous mode. Do not confirm with .<br />
2. Navigate cursor to L/R separate. Do not confirm with .<br />
3. Navigate cursor to an arbitrary pump parameter. Confirm with .<br />
5.5 Switching from univentricular to biventricular operation<br />
Changes:<br />
• The section has been supplemented.<br />
In the change from univentricular to biventricular support, an unintentional change<br />
into the synchronous or separate mode can occur.<br />
In this case:<br />
INSTRUCTION<br />
1. Select the desired mode anew.<br />
2. Check the parameters.<br />