11.14 Error: no data/no reaction from the control processor<br />

Error: no data/no reaction from the control processor<br />

In the event of the simultaneous failure of both control processors or the failure of the<br />

power supply, the <strong>Ikus</strong> cannot generate a specific error message in the message window.<br />

<strong>Instructions</strong><br />

1. Check the status of the patient.<br />

2. Evaluate the failure scenario:<br />

• no alarm, but the <strong>Ikus</strong> continues to function: no communication between the<br />

control processors or between the control processors and the laptop<br />

• optical and acoustic alarm, and the <strong>Ikus</strong> is working in emergency pulse mode: both<br />

control processors are defective<br />

• The <strong>Ikus</strong> is halted; acoustic alarm only: defective power supply<br />

3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />

<strong>Ikus</strong> (see Section 12.4, p. 80).<br />

If no replacement <strong>Ikus</strong> is available and the <strong>Ikus</strong> is running: check the filling and emptying of<br />

the blood pump(s), and monitor the function of the <strong>Ikus</strong>. Do not operate the <strong>Ikus</strong><br />

without supervision!<br />

If no replacement <strong>Ikus</strong> is available and the defective <strong>Ikus</strong> is halted: supply the patient with<br />

the manual pump (see Section 12.5, p. 81)!<br />

4. Always do the following immediately:<br />

Notify hotline! +49(0)30 8187 2772<br />

11.15 The error is only detected by the active (or passive) controller<br />

Only the active (or passive) controller has detected an error. This message only appears<br />

in conjunction with an additional, actual error message. Immediately take all of the<br />

necessary corrective action <strong>for</strong> this second message.<br />

Notify hotline! +49(0)30 8187 2772<br />

11.16 Left/right flow sensor defective. Contact Service department!<br />

The corresponding flow sensor is defective. Although the <strong>Ikus</strong> continues to run, the<br />

excessively low flow would not be detected - possibly due to the snapping off of a cannula<br />

or the driving tube.<br />

Warning!<br />

• Do not operate the <strong>Ikus</strong> without supervision!<br />

<strong>Instructions</strong><br />

1. Check the status of the patient.<br />

2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />

<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />

3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />

<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: supply the patient, if<br />

necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />

manual pump (see Section 12.5, p. 81)!<br />

4. Always do the following immediately:<br />

Notify hotline! +49(0)30 8187 2772<br />

Note<br />

• If the defective <strong>Ikus</strong> is not replaced, this message appears cyclically at 10-minute intervals.<br />

1000002 Rev. 4.9.1 69