Weaning (Bridge-to-Recovery, BTR)<br />

Explantation<br />

10.2 Weaning (Bridge-to-Recovery, BTR)<br />

10.2.1 Explantation with univentricular support<br />

Procedure identical to BTT. Stitch over the insertion sites of all cannulas.<br />

10.2.2 Explantation with biventricular assistance/support<br />

Stop the right pump<br />

<strong>Instructions</strong><br />

1. Select the option “Pa<strong>use</strong> right”, then confirm with . Confirm the decision in the<br />

dialog window with or . The right pump stops. You arrive at the view “Pump size<br />

and single-step mode”. The cursor is on OK.<br />

2. Pull the driving tube of the right pump out of the connector on the <strong>Ikus</strong>. Close the<br />

connector with the seal plug.<br />

3. Confirm with . The <strong>Ikus</strong> continues to work. You arrive at the normal view.<br />

Switch off the <strong>Ikus</strong><br />

<strong>Instructions</strong><br />

1. Supply the patient via the HLM.<br />

2. In the monitoring program, select the option “Drive off”, then confirm with .<br />

3. Confirm the decision in the dialog window with or . The system stops operating<br />

immediately and writes an operating log.<br />

4. Pull the driving tube(s) out of the connector(s). To do so, grab the plug by the ribbed<br />

plug body and pull it out of the connector.<br />

5. Seal the driving tube connection sockets with seal plugs.<br />

6. Wait until the log has been prepared. When the message “Switch off drive at the key<br />

switch” appears, end the monitoring program with , then confirm the decision in<br />

the dialog window with or .<br />

7. In the start menu, select “End” (), then switch off the laptop.<br />

8. When the batteries are completely charged (see below), switch off the <strong>Ikus</strong>. In addition,<br />

set the key switch to OFF [O].<br />

Warning!<br />

• The power switch (mains power switch) on the <strong>Ikus</strong> must always remain on ON/EIN -<br />

even when the <strong>Ikus</strong> is switched off! The only exception is battery mode: Be<strong>for</strong>e switching<br />

to battery mode, move the power switch (mains power switch) to the OFF/Aus [O]<br />

position. Immediately after restoring the power connection, move the power switch back<br />

to the ON/Ein [I] position.<br />

• Always follow the sequence described above. Always switch off the <strong>Ikus</strong> at the key switch.<br />

• Only switch off the <strong>Ikus</strong> when the batteries are completely charged. To do so, leave the<br />

<strong>Ikus</strong> switched on until all the yellow LEDs light up. Then switch off the <strong>Ikus</strong> at the key<br />

switch.<br />

• Always keep un<strong>use</strong>d driving tube connectors closed.<br />

62 1000002 Rev. 4.9.1