Weaning (Bridge-to-Recovery, BTR)<br />
Explantation<br />
10.2 Weaning (Bridge-to-Recovery, BTR)<br />
10.2.1 Explantation with univentricular support<br />
Procedure identical to BTT. Stitch over the insertion sites of all cannulas.<br />
10.2.2 Explantation with biventricular assistance/support<br />
Stop the right pump<br />
<strong>Instructions</strong><br />
1. Select the option “Pa<strong>use</strong> right”, then confirm with . Confirm the decision in the<br />
dialog window with or . The right pump stops. You arrive at the view “Pump size<br />
and single-step mode”. The cursor is on OK.<br />
2. Pull the driving tube of the right pump out of the connector on the <strong>Ikus</strong>. Close the<br />
connector with the seal plug.<br />
3. Confirm with . The <strong>Ikus</strong> continues to work. You arrive at the normal view.<br />
Switch off the <strong>Ikus</strong><br />
<strong>Instructions</strong><br />
1. Supply the patient via the HLM.<br />
2. In the monitoring program, select the option “Drive off”, then confirm with .<br />
3. Confirm the decision in the dialog window with or . The system stops operating<br />
immediately and writes an operating log.<br />
4. Pull the driving tube(s) out of the connector(s). To do so, grab the plug by the ribbed<br />
plug body and pull it out of the connector.<br />
5. Seal the driving tube connection sockets with seal plugs.<br />
6. Wait until the log has been prepared. When the message “Switch off drive at the key<br />
switch” appears, end the monitoring program with , then confirm the decision in<br />
the dialog window with or .<br />
7. In the start menu, select “End” (), then switch off the laptop.<br />
8. When the batteries are completely charged (see below), switch off the <strong>Ikus</strong>. In addition,<br />
set the key switch to OFF [O].<br />
Warning!<br />
• The power switch (mains power switch) on the <strong>Ikus</strong> must always remain on ON/EIN -<br />
even when the <strong>Ikus</strong> is switched off! The only exception is battery mode: Be<strong>for</strong>e switching<br />
to battery mode, move the power switch (mains power switch) to the OFF/Aus [O]<br />
position. Immediately after restoring the power connection, move the power switch back<br />
to the ON/Ein [I] position.<br />
• Always follow the sequence described above. Always switch off the <strong>Ikus</strong> at the key switch.<br />
• Only switch off the <strong>Ikus</strong> when the batteries are completely charged. To do so, leave the<br />
<strong>Ikus</strong> switched on until all the yellow LEDs light up. Then switch off the <strong>Ikus</strong> at the key<br />
switch.<br />
• Always keep un<strong>use</strong>d driving tube connectors closed.<br />
62 1000002 Rev. 4.9.1