Tie-band gun<br />

8.8 Securing the connections<br />

Changes:<br />

• The section has been supplemented.<br />

WARNING<br />

Use only a tie-band gun which has been set correctly. Otherwise the<br />

cannula may be damaged and the patient may lose blood as a result.<br />

Be<strong>for</strong>e using the tie-band gun supplied with the EXCOR tube connecting set, its<br />

settings must be checked and corrected if necessary.<br />

Fig. 18 Tie-band gun<br />

Checking the settings<br />

INSTRUCTION<br />

1. Coarse setting: the letters STD must<br />

be perfectly aligned with the tip of<br />

the arrow embossed on the gun. See<br />

Fig. 19, page 19; 2 and 3.<br />

2. Fine setting: the black bar in the window<br />

must be positioned precisely at<br />

the level of the reference line numbered<br />

5. See Fig. 19, page 19; 1.<br />

1 2 3<br />

Fig. 19 Checking the settings<br />

1000095x14 Rev. 8 19/28