Tie-band gun<br />
8.8 Securing the connections<br />
Changes:<br />
• The section has been supplemented.<br />
WARNING<br />
Use only a tie-band gun which has been set correctly. Otherwise the<br />
cannula may be damaged and the patient may lose blood as a result.<br />
Be<strong>for</strong>e using the tie-band gun supplied with the EXCOR tube connecting set, its<br />
settings must be checked and corrected if necessary.<br />
Fig. 18 Tie-band gun<br />
Checking the settings<br />
INSTRUCTION<br />
1. Coarse setting: the letters STD must<br />
be perfectly aligned with the tip of<br />
the arrow embossed on the gun. See<br />
Fig. 19, page 19; 2 and 3.<br />
2. Fine setting: the black bar in the window<br />
must be positioned precisely at<br />
the level of the reference line numbered<br />
5. See Fig. 19, page 19; 1.<br />
1 2 3<br />
Fig. 19 Checking the settings<br />
1000095x14 Rev. 8 19/28