13.5 Standard <strong>for</strong>m: EXCOR pump log<br />
13.5.1 Explanations on the pump log<br />
7<br />
1 11<br />
2 10<br />
3<br />
5<br />
Numbering of the checkpoints<br />
To briefly describe the findings, we recommend the following letter codes:<br />
p = small speck-like deposits a = small laminar deposits f = small suture t = small thrombus<br />
P = large speck-like deposits A = large laminar deposits F = large suture T = large thrombus<br />
~ = above the respective letter indicates floating deposits<br />
Example: Plotting of the deposits<br />
9<br />
4 8<br />
6<br />
Example: Notation with letter code<br />
Standard <strong>for</strong>m: EXCOR pump log<br />
1 transition inflow cannula - inflow connector<br />
2 Only on pumps with PU valves: Inflow stub in front of inflow valve<br />
3 inflow valve<br />
4 inflow stub behind inflow valve<br />
5 area between inflow and outflow stubs<br />
6 remaining area of blood chamber<br />
7 transition blood chamber - membrane (directly above the rein<strong>for</strong>cement ring)<br />
8 outflow stub in front of outflow valve<br />
9 outflow valve<br />
10 Only on pumps with PU valves: Outflow stub behind outflow valve<br />
11 transition outflow connector - outflow cannula<br />
1 small laminar<br />
3 small suture on the inflow valve<br />
6 small specks<br />
8 larger laminar deposits<br />
1000002 Rev. 4.9.1 93