Do not operate the driving unit without supervision!<br />

Error messages and corrective action<br />

7. Connect the new driving tube to the connector that is now free. The plug snaps audibly<br />

into place. Check that the plug is firmly seated. To do so, pull the plug above the ribbed<br />

plug body. Never pull on the tube!<br />

8. The view “Operating mode” appears in the monitoring program. Select Univentricular or<br />

biventricular, then confirm the desired operating mode with . In biventricular<br />

mode, the view “Pump size and single-step mode” follows. In univentricular mode, the<br />

connector seal test follows first, ad then the view “Pump size and single-step mode”.<br />

9. Confirm with .<br />

10. The system restarts with the set parameters.<br />

11. Observe the pump filling. If necessary, adjust the parameters.<br />

12. Secure the new driving tube on the pump side with cable tie straps.<br />

11.6 Do not operate the driving unit without supervision!<br />

Note<br />

• This message appears if, when the message “Left/right flow sensor defective. Contact<br />

Service department!” on page 69 exists, you do not replace the faulty <strong>Ikus</strong> with an intact<br />

<strong>Ikus</strong>.<br />

Warning!<br />

• Only operate the <strong>Ikus</strong> under supervision!<br />

11.7 Backup operation left/right<br />

The left or right pneumatic system has failed. The corresponding pump is operated by<br />

the backup system.<br />

Notify hotline! +49(0)30 8187 2772<br />

11.8 Please check the tube and the left/right pump<br />

The <strong>Ikus</strong> has detected an excessively high/low flow.<br />

Warning!<br />

• On EXCOR blood pumps with a 10 ml stroke volume, due to the low air volume moved,<br />

the message “Check the left / right tube and pump” is not generated in certain<br />

circumstances. There<strong>for</strong>e, pay particular attention on this pump size to the membrane<br />

movement and the complete filling and emptying of the pump(s).<br />

<strong>Instructions</strong><br />

1. Check the driving tube and cannulas: Is there an obstruction due to bends? Ensure an<br />

unobstructed flow.<br />

2. Check the driving tube and plug: Is the plug position incorrect? Correct the plug<br />

position. Is the driving tube faulty? Replace the driving tube (see Section 11.5.1, p. 65).<br />

3. Adjust the parameters if necessary.<br />

4. Consider changing the cannula position if necessary.<br />

5. Evaluate the hemodynamics of the patient (volume, MAP, PAP, CVP, etc.).<br />

11.9 Fault: ()<br />

Notify hotline! +49(0)30 8187 2772<br />

Note<br />

• If you pass on the error code directly (via phone, fax) to service: Always state all 16 digits!<br />

• Service may request that you read out and send the log files. see Section 12.7 “Read out<br />

the log files” on page 83.<br />

66 1000002 Rev. 4.9.1