11.19 Problem: <br />

If this message appears during the start test: see “Error messages during the start test” on<br />

page 73.<br />

If this message appears outside of the self-test,<br />

a serious problem exists.<br />

The <strong>Ikus</strong> has detected a serious problem.<br />

<strong>Instructions</strong><br />

1. Check the status of the patient.<br />

2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />

<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />

3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />

<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: supply the patient, if<br />

necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />

manual pump (see Section 12.5, p. 81)!<br />

Notify hotline! +49(0)30 8187 2772<br />

11.20 Processor failure. Contact service!<br />

One of the processors has failed. Immediately do the following:<br />

Notify hotline! +49(0)30 8187 2772<br />

11.21 The self-test is not ended by the passive processor<br />

The passive processor was unable to end the self-test of the alarm circuit.<br />

If, within 8 seconds, the message alarm circuit OK appears,<br />

the <strong>Ikus</strong> is working perfectly. No further corrective action is necessary.<br />

If the message alarm circuit OK does not appear within 8 seconds,<br />

the alarm circuit is defective. A possible error might not be detected.<br />

<strong>Instructions</strong><br />

1. Check the status of the patient.<br />

2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />

<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />

3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />

<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: supply the patient, if<br />

necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />

manual pump (see Section 12.5, p. 81)!<br />

4. Always do the following immediately:<br />

Notify hotline! +49(0)30 8187 2772<br />

11.22 System 1 (or 2) is faulty<br />

The corresponding pneumatic system is faulty.<br />

In normal operation<br />

The backup system (system 3) is activated.<br />

Notify hotline! +49(0)30 8187 2772<br />

In backup operation<br />

The <strong>Ikus</strong> is working in emergency operating mode.<br />

Problem: <br />

1000002 Rev. 4.9.1 71