11.19 Problem: <br />
If this message appears during the start test: see “Error messages during the start test” on<br />
page 73.<br />
If this message appears outside of the self-test,<br />
a serious problem exists.<br />
The <strong>Ikus</strong> has detected a serious problem.<br />
<strong>Instructions</strong><br />
1. Check the status of the patient.<br />
2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />
<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />
3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />
<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: supply the patient, if<br />
necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />
manual pump (see Section 12.5, p. 81)!<br />
Notify hotline! +49(0)30 8187 2772<br />
11.20 Processor failure. Contact service!<br />
One of the processors has failed. Immediately do the following:<br />
Notify hotline! +49(0)30 8187 2772<br />
11.21 The self-test is not ended by the passive processor<br />
The passive processor was unable to end the self-test of the alarm circuit.<br />
If, within 8 seconds, the message alarm circuit OK appears,<br />
the <strong>Ikus</strong> is working perfectly. No further corrective action is necessary.<br />
If the message alarm circuit OK does not appear within 8 seconds,<br />
the alarm circuit is defective. A possible error might not be detected.<br />
<strong>Instructions</strong><br />
1. Check the status of the patient.<br />
2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />
<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />
3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />
<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: supply the patient, if<br />
necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />
manual pump (see Section 12.5, p. 81)!<br />
4. Always do the following immediately:<br />
Notify hotline! +49(0)30 8187 2772<br />
11.22 System 1 (or 2) is faulty<br />
The corresponding pneumatic system is faulty.<br />
In normal operation<br />
The backup system (system 3) is activated.<br />
Notify hotline! +49(0)30 8187 2772<br />
In backup operation<br />
The <strong>Ikus</strong> is working in emergency operating mode.<br />
Problem: <br />
1000002 Rev. 4.9.1 71