Failure of the power supply or both control processors<br />

Detecting and eliminating problems<br />

down.<br />

<strong>Instructions</strong><br />

1. The patient should be lying down.<br />

2. Pull out the piston of the manual pump one half stroke length.<br />

3. Remove the driving tube(s) from the <strong>Ikus</strong>. To do so, grab the plug by the ribbed plug<br />

body and pull it out of the connector.<br />

4. Connect the driving tube(s) to the manual pump. Important: Observe the colored<br />

markings.<br />

5. Check that each plug is firmly seated. To do so, pull the plug above the ribbed plug body.<br />

Never pull on the tube!<br />

6. Pump slowly and rhythmically at around 60 to 80 strokes per minute. Important: Move<br />

the piston to its stop, beca<strong>use</strong> this is the only way to exert uni<strong>for</strong>m and controlled<br />

pressures.<br />

7. Check by visual inspection whether the pump is filling and emptying completely.<br />

b<br />

Manual pump<br />

Note<br />

• During biventricular mode: Blood pumps are driven asynchronously by the manual pump.<br />

• Seal the connector(s) of the <strong>Ikus</strong> immediately after removing the driving tube.<br />

12.6 Failure of the power supply or both control processors<br />

In the event of the simultaneous failure of both control processors or the failure of the<br />

power supply, the <strong>Ikus</strong> cannot generate a specific error message in the message window.<br />

An acoustic signal sounds and the visual signal display on the grip lights up. In the<br />

message window, the following message appears: “Error: no data from the control<br />

processor” or “Error: no reaction from the control processor”.<br />

The <strong>Ikus</strong> is working in emergency pulse mode (see Section 12.3, p. 79).<br />

<strong>Instructions</strong><br />

1. Check the status of the patient.<br />

2. Check the filling and emptying of the blood pump(s).<br />

3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />

<strong>Ikus</strong> (see Section 12.4, p. 80).<br />

4. Always do the following immediately:<br />

Notify hotline! +49(0)30 8187 2772<br />

a<br />

a Connector <strong>for</strong> blue marked driving tube<br />

b Connector <strong>for</strong> red marked driving tube<br />

82 1000002 Rev. 4.9.1