Securing the connections<br />

8.8 Securing the connections<br />

Changes:<br />

• The section has been supplemented.<br />

WARNING<br />

Secure each of the connections with 1 cable tie. Otherwise the<br />

connections may loosen and the patient may lose blood as a result.<br />

Visually inspect the cannulae and their connections.<br />

INSTRUCTION<br />

1. Pick up the tube connecting set.<br />

2. Secure each connection between<br />

the cannula or driveline and connector<br />

with a cable tie. To do so, position<br />

the cable tie precisely over the<br />

groove profile of the connector. See<br />

Fig. 16, page 18. IMPORTANT: position<br />

the head of the cable tie so that<br />

it is facing away from the patient's<br />

body.<br />

3. Tighten the cable tie with the tieband<br />

gun. IMPORTANT: observe<br />

section 8.8.<br />

4. Optional: secure the connection with<br />

a second cable tie. 2. Position the<br />

second cable tie above the first one.<br />

The heads of the cable ties must<br />

face in opposite directions and away<br />

from the patient's body. See Fig. 17,<br />

page 18.<br />

5. When using an EXCOR connecting<br />

set: finish by securing each of the<br />

connections with 1 cable tie, follow-<br />

Fig. 16 Correctly positioned cable tie<br />

ing the procedure described above.<br />

Fig. 17 Optional: second cable tie<br />

When using an EXCOR connecting set<br />

INSTRUCTION<br />

1. Finish by securing each of the connections with 1 cable tie, following the<br />

procedure described above.<br />