Diastolic pressure - 100 to 0 mmHg<br />

Precision of pressure display ± 10%<br />

Relative systolic duration 20 to 70%<br />

Off-mains operating time max. 50 min (at max. 30 °C ambient temperature, otherwise less)<br />

Battery charge duration 6 h<br />

Maintenance interval 2000 operating hours or 6 months (at max. 30 °C ambient<br />

temperature, otherwise 1000 operating hours or 3 months)<br />

Max. <strong>Ikus</strong> service life 10 years<br />

13.3 Symbols<br />

13.3.1 <strong>Ikus</strong>: Identification plate<br />

Symbols on the <strong>Ikus</strong> identification plate: Application component type B according to DIN EN<br />

60601-1: 1990 + A1:1993 + A2:1995 (left); certification mark (right)<br />

13.3.2 <strong>Ikus</strong>: connection panel<br />

<strong>Ikus</strong> main switch (key switch) in position OFF [O] (left) and ON [I] (right)<br />

Symbol: Connection of potential equalization according to DIN EN 60601-1: 1990 + A1:1993 + A2:1995<br />

13.3.3 Marking of Numlock status LED<br />

Example <strong>for</strong> marking of the Numlock status LED on the laptop<br />

90 1000002 Rev. 4.9.1