Systole [mmHg]<br />

left/right<br />

Operate the blood pump(s) with the manual pump<br />

2. Select the option “Start program”() from the start menu. Enter the <strong>use</strong>r number and<br />

password, then confirm with . The system runs through the start test.<br />

3. Wait <strong>for</strong> the test phase (a few minutes). Do not acknowledge the acoustic signal.<br />

Messages in the message window in<strong>for</strong>m you about the current status. If the system is<br />

error-free, the “Operating mode” view appears.<br />

4. Select Univentricular or biventricular, then confirm the desired operating mode with<br />

.<br />

5. In biventricular mode, you arrive at the view “Pump size and single-step mode”. In<br />

univentricular mode, the connector seal test follows (duration approx. 10 seconds). The<br />

<strong>Ikus</strong> checks that the blue marked driving tube connector is closed. Then the “Pump size<br />

and single-step mode” view appears.<br />

6. Transfer all parameter values from the old <strong>Ikus</strong> to the replacement <strong>Ikus</strong>.<br />

7. Move the cursor to “Step left”, then confirm with . Biventricular: Move the cursor<br />

to “Step right”, then confirm with .<br />

8. Move the cursor to OK. Important: Do not confirm by clicking OK yet.<br />

Note<br />

• If no parameter values are entered into the replacement <strong>Ikus</strong>, the replacement <strong>Ikus</strong> starts<br />

up with the following default parameter values (standard parameters):<br />

Diastole [mmHg]<br />

left/right<br />

Default standard parameters<br />

Rate [bpm]<br />

• The default settings <strong>for</strong> systole, diastole and relative systolic duration depend on<br />

whether the pump is registered as the left or right pump in the monitoring program.<br />

• Univentricular: Default setting as <strong>for</strong> left pump.<br />

Connecting the blood pump(s) to the replacement <strong>Ikus</strong><br />

Rel. systolic duration [%]<br />

left/right<br />

<strong>Instructions</strong><br />

1. Open the red marked connector (univentricular) or both connectors (biventricular). To<br />

do so, pull the seal plug out of the connector(s).<br />

2. If possible, log on to the monitoring program of the old <strong>Ikus</strong>. Select the option “Drive off”,<br />

then confirm with . Confirm the decision in the dialog window with or .<br />

The <strong>Ikus</strong> sets the operation. If the option “Drive off” cannot be selected, continue with<br />

step 15 without switching off the <strong>Ikus</strong>.<br />

3. As soon as the old <strong>Ikus</strong> has stopped, pull the driving tube(s) off the old <strong>Ikus</strong>. To do so,<br />

grab the plug by the ribbed plug body and pull it out of the connector.<br />

4. Connect the driving tube(s) to the replacement <strong>Ikus</strong>. Important: Observe the colored<br />

markings. Each plug snaps audibly into place. Check that each plug is firmly seated. To<br />

do so, pull the plug above the ribbed plug body. Never pull on the tube!<br />

5. Confirm with . The replacement <strong>Ikus</strong> works with the set parameters.<br />

6. Observe the pump filling. If necessary, adjust the parameters.<br />

7. Switch off the first <strong>Ikus</strong> (see Section 5.3.3, p. 25).<br />

12.5 Operate the blood pump(s) with the manual pump<br />

Operating mode/<br />

mode<br />

210/ 130 -40/ -20 80 40/40 biventricular,<br />

synchronous mode<br />

Necessary if<br />

• the power supply of the <strong>Ikus</strong> is not guaranteed.<br />

• The <strong>Ikus</strong> must be restarted (e.g. in emergency operating mode) and no replacement <strong>Ikus</strong><br />

is available.<br />

Cautionary measure<br />

• Bring in one or more other people to help.<br />

• While supplying the blood pump(s) with the manual pump, your patient should be lying<br />

1000002 Rev. 4.9.1 81