Systole [mmHg]<br />
left/right<br />
Operate the blood pump(s) with the manual pump<br />
2. Select the option “Start program”() from the start menu. Enter the <strong>use</strong>r number and<br />
password, then confirm with . The system runs through the start test.<br />
3. Wait <strong>for</strong> the test phase (a few minutes). Do not acknowledge the acoustic signal.<br />
Messages in the message window in<strong>for</strong>m you about the current status. If the system is<br />
error-free, the “Operating mode” view appears.<br />
4. Select Univentricular or biventricular, then confirm the desired operating mode with<br />
.<br />
5. In biventricular mode, you arrive at the view “Pump size and single-step mode”. In<br />
univentricular mode, the connector seal test follows (duration approx. 10 seconds). The<br />
<strong>Ikus</strong> checks that the blue marked driving tube connector is closed. Then the “Pump size<br />
and single-step mode” view appears.<br />
6. Transfer all parameter values from the old <strong>Ikus</strong> to the replacement <strong>Ikus</strong>.<br />
7. Move the cursor to “Step left”, then confirm with . Biventricular: Move the cursor<br />
to “Step right”, then confirm with .<br />
8. Move the cursor to OK. Important: Do not confirm by clicking OK yet.<br />
Note<br />
• If no parameter values are entered into the replacement <strong>Ikus</strong>, the replacement <strong>Ikus</strong> starts<br />
up with the following default parameter values (standard parameters):<br />
Diastole [mmHg]<br />
left/right<br />
Default standard parameters<br />
Rate [bpm]<br />
• The default settings <strong>for</strong> systole, diastole and relative systolic duration depend on<br />
whether the pump is registered as the left or right pump in the monitoring program.<br />
• Univentricular: Default setting as <strong>for</strong> left pump.<br />
Connecting the blood pump(s) to the replacement <strong>Ikus</strong><br />
Rel. systolic duration [%]<br />
left/right<br />
<strong>Instructions</strong><br />
1. Open the red marked connector (univentricular) or both connectors (biventricular). To<br />
do so, pull the seal plug out of the connector(s).<br />
2. If possible, log on to the monitoring program of the old <strong>Ikus</strong>. Select the option “Drive off”,<br />
then confirm with . Confirm the decision in the dialog window with or .<br />
The <strong>Ikus</strong> sets the operation. If the option “Drive off” cannot be selected, continue with<br />
step 15 without switching off the <strong>Ikus</strong>.<br />
3. As soon as the old <strong>Ikus</strong> has stopped, pull the driving tube(s) off the old <strong>Ikus</strong>. To do so,<br />
grab the plug by the ribbed plug body and pull it out of the connector.<br />
4. Connect the driving tube(s) to the replacement <strong>Ikus</strong>. Important: Observe the colored<br />
markings. Each plug snaps audibly into place. Check that each plug is firmly seated. To<br />
do so, pull the plug above the ribbed plug body. Never pull on the tube!<br />
5. Confirm with . The replacement <strong>Ikus</strong> works with the set parameters.<br />
6. Observe the pump filling. If necessary, adjust the parameters.<br />
7. Switch off the first <strong>Ikus</strong> (see Section 5.3.3, p. 25).<br />
12.5 Operate the blood pump(s) with the manual pump<br />
Operating mode/<br />
mode<br />
210/ 130 -40/ -20 80 40/40 biventricular,<br />
synchronous mode<br />
Necessary if<br />
• the power supply of the <strong>Ikus</strong> is not guaranteed.<br />
• The <strong>Ikus</strong> must be restarted (e.g. in emergency operating mode) and no replacement <strong>Ikus</strong><br />
is available.<br />
Cautionary measure<br />
• Bring in one or more other people to help.<br />
• While supplying the blood pump(s) with the manual pump, your patient should be lying<br />
1000002 Rev. 4.9.1 81