14.3 Electromagnetic immunity - Part 2<br />

Electromagnetic immunity - Part 2<br />

The EXCOR stationary Driving Unit <strong>Ikus</strong> system is intended <strong>for</strong> operation in the electromagnetic<br />

environment described below. The customer or <strong>use</strong>r of the EXCOR stationary<br />

Driving Unit <strong>Ikus</strong> system must ensure that it is <strong>use</strong>d in such an environment.<br />

Noise immunity tests IEC 60601 testing level Compliance level<br />

Derived HF disturbance<br />

variable according to<br />

IEC 61000-4-6<br />

Irradiated HF disturbance<br />

variable according to<br />

IEC 61000-4-3<br />

3 V eff<br />

150 kHz to 80 MHz<br />

outside of the ISM<br />

bands a<br />

10 V eff<br />

150 kHz to 80 MHz<br />

within the ISM bands a<br />

10 V/m<br />

80 MHz to 2.5 GHz<br />

[V 1]=10 V<br />

[V 2]=10 V<br />

[E 1 ]=10 V/m<br />

Electromagnetic environment -<br />

guidelines<br />

Wearable and mobile radio devices<br />

should not be <strong>use</strong>d at a shorter distance<br />

from the EXCOR stationary Driving Unit<br />

<strong>Ikus</strong> system including its cables than the<br />

recommended protection distance<br />

calculated according to the appropriate<br />

equation <strong>for</strong> the transmission<br />

frequencies.<br />

Recommended protection distance<br />

d = 0,<br />

35<br />

d = 1,<br />

2<br />

d = 1,<br />

2<br />

d =<br />

2,<br />

3<br />

80 MHz to 800 MHz<br />

800 MHz to 2.5 GHz<br />

With P as the maximum rated power<br />

of the transmitter in watts (W)<br />

according to the data from the<br />

transmitter manufacturer and d as the<br />

recommended protection distance in<br />

meters (m). b The field strength of<br />

stationary radio transmitters should,<br />

at all frequencies according to an<br />

on-site investigation, c be less than<br />

the compliance level. d Interference is<br />

possible in the vicinity of devices that<br />

bear the following pictogram.<br />

• NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.<br />

• NOTE 2 These guidelines may not be applicable to all cases. The propagation of electromagnetic variables is influenced by<br />

absorption and reflections from the building, objects and people.<br />

• a) The ISM frequency bands (<strong>for</strong> industrial, scientific and medical applications) between 150 kHz and 80MHz are 6,765 MHz<br />

to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz and 40,66MHz to 40,70MHz.<br />

• b) The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range of 80 MHz to<br />

2.5 GHz are set in order to reduce the probability of mobile/wearable communication devices causing interference if they<br />

are inadvertently brought into the vicinity of the patient. For this reason, the additional factor of 10/3 is applied when calculating<br />

the recommended protection distances in these frequency ranges.<br />

• c) The field strength of stationary transmitters, e.g. base stations <strong>for</strong> cellular telephones and mobile terrestrial radio devices,<br />

amateur radio stations, AM and FM radio and television transmitters can theoretically not be predetermined precisely.<br />

To determine the electromagnetic environment with respect to the stationary transmitter, a study of the site should be considered.<br />

If the measured field strength at the location exeeds the above con<strong>for</strong>mance level, the normal operation <strong>for</strong> the<br />

EXCOR stationary Driving Unit <strong>Ikus</strong> system must be monitored at each site of <strong>use</strong> to demonstrate proper function. If unusual<br />

per<strong>for</strong>mance features are observed, additional corrective action may be required, e.g. a modified orientation or another<br />

site <strong>for</strong> the EXCOR stationary Driving Unit <strong>Ikus</strong> system .<br />

• d) Over the frequency range 150 kHz to 80 MHz, the field strength should be less than 10 V/m.<br />

1000002 Rev. 4.9.1 103<br />

P<br />

P<br />

P<br />

P