14.3 Electromagnetic immunity - Part 2<br />
Electromagnetic immunity - Part 2<br />
The EXCOR stationary Driving Unit <strong>Ikus</strong> system is intended <strong>for</strong> operation in the electromagnetic<br />
environment described below. The customer or <strong>use</strong>r of the EXCOR stationary<br />
Driving Unit <strong>Ikus</strong> system must ensure that it is <strong>use</strong>d in such an environment.<br />
Noise immunity tests IEC 60601 testing level Compliance level<br />
Derived HF disturbance<br />
variable according to<br />
IEC 61000-4-6<br />
Irradiated HF disturbance<br />
variable according to<br />
IEC 61000-4-3<br />
3 V eff<br />
150 kHz to 80 MHz<br />
outside of the ISM<br />
bands a<br />
10 V eff<br />
150 kHz to 80 MHz<br />
within the ISM bands a<br />
10 V/m<br />
80 MHz to 2.5 GHz<br />
[V 1]=10 V<br />
[V 2]=10 V<br />
[E 1 ]=10 V/m<br />
Electromagnetic environment -<br />
guidelines<br />
Wearable and mobile radio devices<br />
should not be <strong>use</strong>d at a shorter distance<br />
from the EXCOR stationary Driving Unit<br />
<strong>Ikus</strong> system including its cables than the<br />
recommended protection distance<br />
calculated according to the appropriate<br />
equation <strong>for</strong> the transmission<br />
frequencies.<br />
Recommended protection distance<br />
d = 0,<br />
35<br />
d = 1,<br />
2<br />
d = 1,<br />
2<br />
d =<br />
2,<br />
3<br />
80 MHz to 800 MHz<br />
800 MHz to 2.5 GHz<br />
With P as the maximum rated power<br />
of the transmitter in watts (W)<br />
according to the data from the<br />
transmitter manufacturer and d as the<br />
recommended protection distance in<br />
meters (m). b The field strength of<br />
stationary radio transmitters should,<br />
at all frequencies according to an<br />
on-site investigation, c be less than<br />
the compliance level. d Interference is<br />
possible in the vicinity of devices that<br />
bear the following pictogram.<br />
• NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.<br />
• NOTE 2 These guidelines may not be applicable to all cases. The propagation of electromagnetic variables is influenced by<br />
absorption and reflections from the building, objects and people.<br />
• a) The ISM frequency bands (<strong>for</strong> industrial, scientific and medical applications) between 150 kHz and 80MHz are 6,765 MHz<br />
to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz and 40,66MHz to 40,70MHz.<br />
• b) The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range of 80 MHz to<br />
2.5 GHz are set in order to reduce the probability of mobile/wearable communication devices causing interference if they<br />
are inadvertently brought into the vicinity of the patient. For this reason, the additional factor of 10/3 is applied when calculating<br />
the recommended protection distances in these frequency ranges.<br />
• c) The field strength of stationary transmitters, e.g. base stations <strong>for</strong> cellular telephones and mobile terrestrial radio devices,<br />
amateur radio stations, AM and FM radio and television transmitters can theoretically not be predetermined precisely.<br />
To determine the electromagnetic environment with respect to the stationary transmitter, a study of the site should be considered.<br />
If the measured field strength at the location exeeds the above con<strong>for</strong>mance level, the normal operation <strong>for</strong> the<br />
EXCOR stationary Driving Unit <strong>Ikus</strong> system must be monitored at each site of <strong>use</strong> to demonstrate proper function. If unusual<br />
per<strong>for</strong>mance features are observed, additional corrective action may be required, e.g. a modified orientation or another<br />
site <strong>for</strong> the EXCOR stationary Driving Unit <strong>Ikus</strong> system .<br />
• d) Over the frequency range 150 kHz to 80 MHz, the field strength should be less than 10 V/m.<br />