6 Start up the <strong>Ikus</strong> and set the parameters<br />
Preparations outside of the operating room<br />
Warning!<br />
• Be<strong>for</strong>e every commissioning of the <strong>Ikus</strong>, you must check that a replacement <strong>Ikus</strong> or<br />
<strong>Berlin</strong> <strong>Heart</strong> GmbH service is available within 24 hours.<br />
• If you are using a laptop of the type Panasonic CF29, always observe the additional<br />
in<strong>for</strong>mation (Section 13.9 on p. 99)!<br />
Caution!<br />
• Always observe a resting period of six hours after each transportation be<strong>for</strong>e switching<br />
on the <strong>Ikus</strong>.<br />
• Be<strong>for</strong>e every commissioning of the <strong>Ikus</strong>, check that the environmental conditions are<br />
respected (see Section 13.2 “Technical specifications” on page 89).<br />
• Always keep un<strong>use</strong>d driving tube connectors closed.<br />
Note<br />
• This chapter describes the technical aspects of comissioning as well as perioperative<br />
and postoperative drive management. During commissioning, always pay attention to<br />
the chapter 8: “Implantation – Surgical procedure”, p. 45.<br />
• The described procedure <strong>for</strong> drive management consists of recommendations. They do<br />
not replace conscientious observation and evaluation by competent personnel.<br />
6.1 Preparations outside of the operating room<br />
•<br />
Warning!<br />
Switch on the <strong>Ikus</strong> two hours be<strong>for</strong>e its planned <strong>use</strong>, so as to charge the batteries<br />
(in particular the laptop battery) sufficiently and to detect a possible device error by<br />
means of the start test. During this period, always connect the vessel models to the <strong>Ikus</strong><br />
(see below).<br />
• The power switch (mains power switch) on the <strong>Ikus</strong> must always remain on ON/EIN –<br />
even in battery mode and when the <strong>Ikus</strong> is switched off! Otherwise, there is a risk of<br />
failure of the driving unit due to completely discharged batteries. The only exception is<br />
battery mode: Be<strong>for</strong>e switching to battery mode, move the power switch (mains power<br />
switch) to the OFF/Aus [O] position. Immediately after restoring the power connection,<br />
move the power switch back to the ON/Ein [I] position.<br />
• Make sure that the key on the laptop is not activated. The status LED on the<br />
laptop indicated with a lock and/or a number (e.g. “1”) should not light up.<br />
1 2 3<br />
8<br />
Connection panel<br />
7 6 5<br />
1000002 Rev. 4.9.1 31<br />
4<br />
4<br />
1. Driving tube connectors<br />
2. Battery f<strong>use</strong><br />
3. Mains power indicator<br />
4. Power switch (mains power switch)<br />
5. Mains power connector<br />
6. Main switch (key switch; O/I)<br />
7. Potential equalization connection<br />
8. External alarm connector (nurse call)