FATAL: Battery failure! In<strong>for</strong>m Service urgently!<br />
Error messages and corrective action<br />
<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: Supply the patient, if<br />
necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />
manual pump (see Section 12.5, p. 81)!<br />
4. Always do the following immediately:<br />
Notify hotline! +49(0)30 8187 2772<br />
11.12.3 ...is reporting the error: Flow measurement on the left (or right)<br />
The backup processor has detected a flow error, but the main processor has not detected<br />
a flow error.<br />
<strong>Instructions</strong><br />
1. Check the filling and emptying of the blood pump(s).<br />
2. Check the plausibility of the measured values.<br />
3. Reset all of the parameter values. In addition, log out of the monitoring program and then<br />
log back into it. Navigate the cursor with ← / → to the desired field, adjust the value with<br />
↓,↑/ / , then confirm with . The system works with the new value.<br />
If the message appears again:<br />
Notify hotline! +49(0)30 8187 2772<br />
11.12.4 ...is reporting a checking error<br />
The main processor cannot end the test phase to check the system after the error<br />
message “Pressure error/time error in system 1 (or 2 or 3)” (see Section 11.11, p. 67).<br />
Warning!<br />
• The blood pump driven by the tested system has been stopped since the start of the test<br />
phase. Take corrective action immediately!<br />
<strong>Instructions</strong><br />
1. Evaluate the status of the patient.<br />
2. Immediately ensure the supply of the patient by means of a replacement <strong>Ikus</strong> (see<br />
Section 12.4, p. 80) or a manual pump (see Section 12.5, p. 81).<br />
Notify hotline! +49(0)30 8187 2772<br />
11.13 FATAL: Battery failure! In<strong>for</strong>m Service urgently!<br />
Warning!<br />
• You must continue to operate the <strong>Ikus</strong> on mains power! There is a threat of total failure<br />
if battery mode is <strong>use</strong>d!<br />
<strong>Instructions</strong><br />
1. In the case of battery mode: Connect the <strong>Ikus</strong> to the mains power supply immediately!<br />
2. Check the status of the patient.<br />
3. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />
<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />
4. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />
<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: Supply the patient, if<br />
necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />
manual pump (see Section 12.5, p. 81)!<br />
5. Always do the following immediately:<br />
Notify hotline! +49(0)30 8187 2772<br />
68 1000002 Rev. 4.9.1