FATAL: Battery failure! In<strong>for</strong>m Service urgently!<br />

Error messages and corrective action<br />

<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: Supply the patient, if<br />

necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />

manual pump (see Section 12.5, p. 81)!<br />

4. Always do the following immediately:<br />

Notify hotline! +49(0)30 8187 2772<br />

11.12.3 ...is reporting the error: Flow measurement on the left (or right)<br />

The backup processor has detected a flow error, but the main processor has not detected<br />

a flow error.<br />

<strong>Instructions</strong><br />

1. Check the filling and emptying of the blood pump(s).<br />

2. Check the plausibility of the measured values.<br />

3. Reset all of the parameter values. In addition, log out of the monitoring program and then<br />

log back into it. Navigate the cursor with ← / → to the desired field, adjust the value with<br />

↓,↑/ / , then confirm with . The system works with the new value.<br />

If the message appears again:<br />

Notify hotline! +49(0)30 8187 2772<br />

11.12.4 ...is reporting a checking error<br />

The main processor cannot end the test phase to check the system after the error<br />

message “Pressure error/time error in system 1 (or 2 or 3)” (see Section 11.11, p. 67).<br />

Warning!<br />

• The blood pump driven by the tested system has been stopped since the start of the test<br />

phase. Take corrective action immediately!<br />

<strong>Instructions</strong><br />

1. Evaluate the status of the patient.<br />

2. Immediately ensure the supply of the patient by means of a replacement <strong>Ikus</strong> (see<br />

Section 12.4, p. 80) or a manual pump (see Section 12.5, p. 81).<br />

Notify hotline! +49(0)30 8187 2772<br />

11.13 FATAL: Battery failure! In<strong>for</strong>m Service urgently!<br />

Warning!<br />

• You must continue to operate the <strong>Ikus</strong> on mains power! There is a threat of total failure<br />

if battery mode is <strong>use</strong>d!<br />

<strong>Instructions</strong><br />

1. In the case of battery mode: Connect the <strong>Ikus</strong> to the mains power supply immediately!<br />

2. Check the status of the patient.<br />

3. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />

<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />

4. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />

<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: Supply the patient, if<br />

necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />

manual pump (see Section 12.5, p. 81)!<br />

5. Always do the following immediately:<br />

Notify hotline! +49(0)30 8187 2772<br />

68 1000002 Rev. 4.9.1