Restart the <strong>Ikus</strong><br />
Detecting and eliminating problems<br />
Never pull on the tube!<br />
14. Loosen the tube clamps on the cannulas.<br />
15. In the monitoring program, move the cursor to the single-step mode field. Confirm<br />
“Step left” with to trigger a single-step. With , trigger additional<br />
individual steps until the left pump is completely de-aired.<br />
16. Move the cursor to “Step right”, then confirm with to trigger a single-step. With<br />
, trigger additional individual steps until the right pump is completely de-aired.<br />
17. If both pumps are completely de-aired: Remove the de-airing needles.<br />
18. Move the cursor to OK, then confirm with . The system restarts with the set<br />
parameters.<br />
19. Observe the pump filling. If necessary, adjust the parameters.<br />
20. Secure all the connections with cable tie straps.<br />
12.2 Restart the <strong>Ikus</strong><br />
Warning!<br />
• Only switch off and restart the <strong>Ikus</strong> when requested to do so by service (e.g. in emergency<br />
operating mode).<br />
• The power switch (mains power switch) on the <strong>Ikus</strong> must always remain on ON/EIN -<br />
even when the <strong>Ikus</strong> is switched off! The only exception is battery mode: Be<strong>for</strong>e switching<br />
to battery mode, move the power switch (mains power switch) to the OFF/Aus [O]<br />
position. Immediately after restoring the power connection, move the power switch back<br />
to the ON/Ein [I] position.<br />
• If the curves in the monitoring program are at a standstill and the parameters cannot be<br />
adjusted even after restarting the laptop and monitoring program, the <strong>Ikus</strong> works in<br />
emergency pulse mode. In this case, do not proceed as described here. Proceed<br />
according to section 12.3, “Emergency pulse mode”, p. 79.<br />
Note<br />
• If the <strong>Ikus</strong> is only restarted after a controlled shutdown, the <strong>Ikus</strong> runs through the start<br />
test and all of its parameters must be set again (see page 81).<br />
• If the <strong>Ikus</strong> is inadvertently switched off at the key switch and immediately restarted, the<br />
<strong>Ikus</strong> resumes working with the last set parameters. If the <strong>Ikus</strong> is only restarted after more<br />
than five minutes, the <strong>Ikus</strong> runs through the start test and all of its parameters must be<br />
set again. If the parameters are not set again, the <strong>Ikus</strong> works with the default standard<br />
parameters (see page 81).<br />
<strong>Instructions</strong><br />
1. Supply the patient with a replacement <strong>Ikus</strong> (see Section 12.4, p. 80) or manual pump<br />
(see Section 12.5, p. 81).<br />
2. Seal the driving tube connectors on the <strong>Ikus</strong> with seal plugs.<br />
3. Switch off the <strong>Ikus</strong>. To do so, in the monitoring program, select the option “Drive off”,<br />
then confirm with . Confirm the decision in the dialog window with or .<br />
Important: If the option “Drive off” cannot be selected, the <strong>Ikus</strong> is working in emergency<br />
pulse mode. In this case, proceed as described in the section 12.3, “Emergency pulse<br />
mode”, p. 79.<br />
4. Wait until the log has been prepared. In the case of the message “Switch off drive at key<br />
switch”, set the key switch to OFF [O].<br />
5. Switch off the laptop.<br />
6. Switch on the <strong>Ikus</strong> again immediately. To do so, set the key switch to ON [I].<br />
7. Switch on the laptop. Select the option “Start program”() from the start menu. Enter<br />
the <strong>use</strong>r number and password, then confirm with .<br />
8. Check all of the parameters and adjust them if necessary.<br />
9. Open the red marked connector (univentricular) or both connectors (biventricular). To<br />
do so, pull the seal plug out of the connector(s).<br />
10. Pull the driving tube(s) off the replacement <strong>Ikus</strong> or manual pump and connect them to<br />
the <strong>Ikus</strong>. Important: Observe the colored markings. Each plug snaps audibly into place.<br />
11. Check that each plug is firmly seated. To do so, pull the plug above the ribbed plug body.<br />