Restart the <strong>Ikus</strong><br />

Detecting and eliminating problems<br />

Never pull on the tube!<br />

14. Loosen the tube clamps on the cannulas.<br />

15. In the monitoring program, move the cursor to the single-step mode field. Confirm<br />

“Step left” with to trigger a single-step. With , trigger additional<br />

individual steps until the left pump is completely de-aired.<br />

16. Move the cursor to “Step right”, then confirm with to trigger a single-step. With<br />

, trigger additional individual steps until the right pump is completely de-aired.<br />

17. If both pumps are completely de-aired: Remove the de-airing needles.<br />

18. Move the cursor to OK, then confirm with . The system restarts with the set<br />

parameters.<br />

19. Observe the pump filling. If necessary, adjust the parameters.<br />

20. Secure all the connections with cable tie straps.<br />

12.2 Restart the <strong>Ikus</strong><br />

Warning!<br />

• Only switch off and restart the <strong>Ikus</strong> when requested to do so by service (e.g. in emergency<br />

operating mode).<br />

• The power switch (mains power switch) on the <strong>Ikus</strong> must always remain on ON/EIN -<br />

even when the <strong>Ikus</strong> is switched off! The only exception is battery mode: Be<strong>for</strong>e switching<br />

to battery mode, move the power switch (mains power switch) to the OFF/Aus [O]<br />

position. Immediately after restoring the power connection, move the power switch back<br />

to the ON/Ein [I] position.<br />

• If the curves in the monitoring program are at a standstill and the parameters cannot be<br />

adjusted even after restarting the laptop and monitoring program, the <strong>Ikus</strong> works in<br />

emergency pulse mode. In this case, do not proceed as described here. Proceed<br />

according to section 12.3, “Emergency pulse mode”, p. 79.<br />

Note<br />

• If the <strong>Ikus</strong> is only restarted after a controlled shutdown, the <strong>Ikus</strong> runs through the start<br />

test and all of its parameters must be set again (see page 81).<br />

• If the <strong>Ikus</strong> is inadvertently switched off at the key switch and immediately restarted, the<br />

<strong>Ikus</strong> resumes working with the last set parameters. If the <strong>Ikus</strong> is only restarted after more<br />

than five minutes, the <strong>Ikus</strong> runs through the start test and all of its parameters must be<br />

set again. If the parameters are not set again, the <strong>Ikus</strong> works with the default standard<br />

parameters (see page 81).<br />

<strong>Instructions</strong><br />

1. Supply the patient with a replacement <strong>Ikus</strong> (see Section 12.4, p. 80) or manual pump<br />

(see Section 12.5, p. 81).<br />

2. Seal the driving tube connectors on the <strong>Ikus</strong> with seal plugs.<br />

3. Switch off the <strong>Ikus</strong>. To do so, in the monitoring program, select the option “Drive off”,<br />

then confirm with . Confirm the decision in the dialog window with or .<br />

Important: If the option “Drive off” cannot be selected, the <strong>Ikus</strong> is working in emergency<br />

pulse mode. In this case, proceed as described in the section 12.3, “Emergency pulse<br />

mode”, p. 79.<br />

4. Wait until the log has been prepared. In the case of the message “Switch off drive at key<br />

switch”, set the key switch to OFF [O].<br />

5. Switch off the laptop.<br />

6. Switch on the <strong>Ikus</strong> again immediately. To do so, set the key switch to ON [I].<br />

7. Switch on the laptop. Select the option “Start program”() from the start menu. Enter<br />

the <strong>use</strong>r number and password, then confirm with .<br />

8. Check all of the parameters and adjust them if necessary.<br />

9. Open the red marked connector (univentricular) or both connectors (biventricular). To<br />

do so, pull the seal plug out of the connector(s).<br />

10. Pull the driving tube(s) off the replacement <strong>Ikus</strong> or manual pump and connect them to<br />

the <strong>Ikus</strong>. Important: Observe the colored markings. Each plug snaps audibly into place.<br />

11. Check that each plug is firmly seated. To do so, pull the plug above the ribbed plug body.<br />

78 1000002 Rev. 4.9.1