13.6 Standard <strong>for</strong>m: EXCOR parameter log<br />

Patient: Lot no. of left pump: Lot no. of right pump:<br />

Standard <strong>for</strong>m: EXCOR parameter log<br />

Left pump vol. Rate Drive pressure Membrane movement<br />

Right pump vol. left right left (mmHg) right (mmHg) left right<br />

Date Time Abbreviation bpm bpm Systole Diastole Systole Diastole Emptying Filling Emptying Filling<br />

Example 8:00 e.g. 60 60 180 - 30 130 - 30 + + + - ++ ++<br />

Membrane movement: Explanation of the notation<br />

• + + : Complete or almost complete filling (max. 1 small fold in the membrane), membrane moves<br />

to the end-of-systole or end-of-diastole position<br />

• + - : incomplete filling or emptying movement (membrane shows more than one fold, if it is in the<br />

end-of-systole or end-of-diastole position)<br />

• - - : incorrect filling or emptying (membrane moves only partially or not at all)<br />

1000002 Rev. 4.9.1 95