13.6 Standard <strong>for</strong>m: EXCOR parameter log<br />
Patient: Lot no. of left pump: Lot no. of right pump:<br />
Standard <strong>for</strong>m: EXCOR parameter log<br />
Left pump vol. Rate Drive pressure Membrane movement<br />
Right pump vol. left right left (mmHg) right (mmHg) left right<br />
Date Time Abbreviation bpm bpm Systole Diastole Systole Diastole Emptying Filling Emptying Filling<br />
Example 8:00 e.g. 60 60 180 - 30 130 - 30 + + + - ++ ++<br />
Membrane movement: Explanation of the notation<br />
• + + : Complete or almost complete filling (max. 1 small fold in the membrane), membrane moves<br />
to the end-of-systole or end-of-diastole position<br />
• + - : incomplete filling or emptying movement (membrane shows more than one fold, if it is in the<br />
end-of-systole or end-of-diastole position)<br />
• - - : incorrect filling or emptying (membrane moves only partially or not at all)<br />
1000002 Rev. 4.9.1 95