Connect the pumps to the cannulas<br />

Implantation – Surgical procedure<br />

8.6 Connect the pumps to the cannulas<br />

Warning!<br />

• When connecting the blood pump(s), pay attention to the arrows on the inflow and<br />

outflow stubs. These show the blood flow direction.<br />

Note<br />

• Depending on the site of the anastomosis, the air or blood chambers of the pump<br />

point upward:<br />

Support<br />

LVAD<br />

but<br />

LVAD with P80M-004 and P80M-005<br />

(in/ out exchanged)<br />

Anastomosis of inflow<br />

cannula in<br />

points upward...<br />

<strong>Instructions</strong><br />

1. Bring the patient into the Trendelenburg position.<br />

2. Loosen the tube clamps, rinse the cannulas, then lock the cannulas underneath the<br />

incision wound with tube clamps.<br />

3. Connect the prefilled pump first on the inflow site and then on the outflow side. In doing<br />

so, sprinkle sterile saline solution on it to connect the pump approximately without air.<br />

Proceed carefully to prevent damage to the glove or the inner surface of the cannula.<br />

4. Loosen the tube clamps, then de-air the pump(s) and cannulas.<br />

5. Connect the driving tube to the pump.<br />

Caution!<br />

• Ensure the tension-free and bend-free positioning of the cannulas, blood pump(s) and<br />

driving tubes.<br />

End position of the blood pumps with biventricular assistance/ support, left apical<br />

Note<br />

• Then the driving tube is connected to the <strong>Ikus</strong>. The <strong>Ikus</strong> is started and the parameters<br />

50 1000002 Rev. 4.9.1<br />

Apex<br />

Apex<br />

Blood chamber<br />

Air chamber<br />

RVAD Atrium Air chamber<br />

LVAD Atrium Air chamber<br />

a<br />

b<br />

c<br />

f<br />

e<br />

d<br />

a RVAD/ right pump (air chamber points upward)<br />

b Inflow cannula of the right atrium<br />

c Outflow cannula to the pulmonary artery<br />

d LVAD/ left pump (blood chamber points upward)<br />

e Inflow cannula of the apex<br />

f Outflow cannula to the aorta ascendens