Routine start test in the event of non-<strong>use</strong><br />
<strong>Instructions</strong> <strong>for</strong> <strong>use</strong>: <strong>Ikus</strong><br />
snaps audibly into place. Check that the plug is firmly seated. To do so, pull the plug<br />
above the ribbed plug body. Never pull on the tube!<br />
6. Select the operating mode biventricular, then confirm with . All additional steps<br />
are analogous to the start procedure (see page 33 ff.).<br />
5.6 Routine start test in the event of non-<strong>use</strong><br />
Caution!<br />
• When the <strong>Ikus</strong> is not <strong>use</strong>d, start it up <strong>for</strong> 24 hours each month to ensure that all the<br />
batteries (especially the laptop batteries) are sufficiently charged, and per<strong>for</strong>m the<br />
start test.<br />
<strong>Instructions</strong><br />
1. Procedure as described in the section 6.1.1: “Connecting the vessel model”, p. 32 and<br />
6.1.2: “Switch on the <strong>Ikus</strong>”, p. 32.<br />
2. Wait 24 hours.<br />
3. Procedure as described in 6.1.3: “Starting the monitoring program”, p. 32 and 6.1.4:<br />
“Setting the test parameters”, p. 33 up to procedure step 2.<br />
4. Observe and evaluate the curve representation.<br />
5. Select the option “Drive off”, then confirm with .<br />
6. Confirm the decision in the dialog window with or . The system sets the<br />
operation immediately and writes an operating log.<br />
7. Wait until the log has been prepared (message: “Switch off the drive at the key switch”).<br />
8. End the monitoring program and switch off the laptop.<br />
9. Provided that all of the yellow battery LEDs are illuminated, switch off the <strong>Ikus</strong>.<br />
In addition, set the key switch to OFF [O].<br />
10. Disconnect both pressure vessels from the <strong>Ikus</strong>.<br />
11. Seal the driving tube connection sockets with seal plugs.<br />
If the system detects an error in the start test<br />
<strong>Instructions</strong><br />
1. End the monitoring program according to the instructions in the dialog.<br />
2. Switch off the <strong>Ikus</strong> at the key switch. Important: The power switch (mains power switch)<br />
remains on ON/EIN!<br />
3. Important: Wait 10 minutes.<br />
4. Restart the start procedure (see above).<br />
If the system again detects an error in the start test and/or not all of the battery LEDs are<br />
illuminated after 24 hours:<br />
Notify hotline! +49(0)30 8187 2772<br />
28 1000002 Rev. 4.9.1