6.2.4 Setting the startup parameters<br />

Systole [mmHg]<br />

left/right<br />

“Pump size and single-step mode” view with recommended startup parameters<br />

Warning!<br />

• Especially after previous ECMO it is necessary to change the settings of the start<br />

parameters be<strong>for</strong>e starting the system.<br />

Intraoperative drive management<br />

<strong>Instructions</strong><br />

1. Place the cursor with ↑ in the parameter table and navigate with ← / → to the desired field<br />

in the parameter table. Adjust the value with ↓,↑ or / , then confirm with<br />

.<br />

2. Setting the startup parameters:<br />

Startup parameters<br />

Diastole [mmHg]<br />

left/right<br />

6.2.5 Connect the blood pump(s) to the <strong>Ikus</strong><br />

Rate [bpm]<br />

100/ 80 -5/-5 30 40/40<br />

Note<br />

• Condition of the blood pump(s) on connecting: Filled with sterile saline solution, with<br />

the de-airing needle fitted. For easier handling after connecting the pump(s), the driving<br />

tubes are not connected until after the inflow and outflow cannulae are connected to the<br />

pump (see page 50).<br />

<strong>Instructions</strong><br />

1. Open the red marked connector (univentricular) or both connectors (biventricular).<br />

To do so, pull the seal plug out of the connector(s).<br />

2. Connect the driving tube to the <strong>Ikus</strong>. To do so, insert the plug of the tube into the<br />

connector. The plug snaps audibly into place. Check that each plug is firmly seated.<br />

To do so, pull the plug above the ribbed plug body. Never pull on the tube!<br />

3. During biventricular mode: Observe the colored markings.<br />

4. During biventricular mode: Repeat with the second pump.<br />

Operating mode Connector on the <strong>Ikus</strong><br />

biventricular LVAD: red marked connector<br />

RVAD: blue marked connector<br />

univentricular LVAD or RVAD: red marked connector<br />

Allocation: Operating mode, blood pump, connector<br />

Rel. systolic duration [%]<br />

left/right<br />

1000002 Rev. 4.9.1 35