6.2.4 Setting the startup parameters<br />
Systole [mmHg]<br />
left/right<br />
“Pump size and single-step mode” view with recommended startup parameters<br />
Warning!<br />
• Especially after previous ECMO it is necessary to change the settings of the start<br />
parameters be<strong>for</strong>e starting the system.<br />
Intraoperative drive management<br />
<strong>Instructions</strong><br />
1. Place the cursor with ↑ in the parameter table and navigate with ← / → to the desired field<br />
in the parameter table. Adjust the value with ↓,↑ or / , then confirm with<br />
.<br />
2. Setting the startup parameters:<br />
Startup parameters<br />
Diastole [mmHg]<br />
left/right<br />
6.2.5 Connect the blood pump(s) to the <strong>Ikus</strong><br />
Rate [bpm]<br />
100/ 80 -5/-5 30 40/40<br />
Note<br />
• Condition of the blood pump(s) on connecting: Filled with sterile saline solution, with<br />
the de-airing needle fitted. For easier handling after connecting the pump(s), the driving<br />
tubes are not connected until after the inflow and outflow cannulae are connected to the<br />
pump (see page 50).<br />
<strong>Instructions</strong><br />
1. Open the red marked connector (univentricular) or both connectors (biventricular).<br />
To do so, pull the seal plug out of the connector(s).<br />
2. Connect the driving tube to the <strong>Ikus</strong>. To do so, insert the plug of the tube into the<br />
connector. The plug snaps audibly into place. Check that each plug is firmly seated.<br />
To do so, pull the plug above the ribbed plug body. Never pull on the tube!<br />
3. During biventricular mode: Observe the colored markings.<br />
4. During biventricular mode: Repeat with the second pump.<br />
Operating mode Connector on the <strong>Ikus</strong><br />
biventricular LVAD: red marked connector<br />
RVAD: blue marked connector<br />
univentricular LVAD or RVAD: red marked connector<br />
Allocation: Operating mode, blood pump, connector<br />
Rel. systolic duration [%]<br />
left/right<br />
1000002 Rev. 4.9.1 35