Diagnostic functions<br />

Low Flow PV-Loop (optional)<br />

Low Flow PV-Loop (optional)<br />

<strong>EvitaXL</strong> determines the Low Flow<br />

PV-Loop* during an extended inspiration<br />

or an inspiration and expiration.<br />

The measuring procedure can only be<br />

carried out in the »Adult« patient mode.<br />

The measuring procedure should only<br />

be carried out on patients with no spontaneous<br />

breathing.<br />

● Touch the function key »Special Procedure...«.<br />

<strong>EvitaXL</strong> displays the menu<br />

»Additional Function«.<br />

● Touch the screen key<br />

»Low Flow PV-Loop«.<br />

<strong>EvitaXL</strong> opens the information screen.<br />

Observe the information before carrying<br />

out the measurement!<br />

For additional information, see page 227.<br />

* Lung Protection Package option<br />

For a detailed description, please refer to<br />

page 227.<br />

The application of a low flow manoeuvre may decrease the patient's systemic circulatory<br />

pressure and could cause a pneumothorax, for example. Carefully assess the<br />

patients's condition for settings.<br />

— Applied pressure / volume must be adequate for the patient.<br />

— The patient must be haemodynamically stable.<br />

— Closely monitor arterial blood pressure during the manoeuvre.<br />

— The sudden release of high airway pressure may overload the heart and impair<br />

cardiac functions.<br />

— The calculated maximum manoeuvre time must be adequate for the patient.<br />

— These measurements are only valid with no spontaneous breathing.<br />

— These measurements are only valid with no leakage. Volume and derived<br />

compliance values are not leakage compensated (VTi, VTe and Cstat).<br />

— This manoeuvre cannot be restarted within 60 seconds.<br />

— This manoeuvre cannot be started until 60 seconds after nebulisation or<br />

suctioning.<br />