Technical Data<br />

Breathing Support Package (optional)<br />

O2 Therapy (optional)<br />

Breathing Support Package (optional)<br />

Adult mode<br />

Settings for PPS:<br />

Flow Assist (FlowAssist)<br />

Range 0 to 30 mbar/L/s<br />

Resolution 0.5 mbar/L/s<br />

Corresponds to resistance compensation 0 to 30 mbar/L/s<br />

Volume Assist (Vol.Assist)<br />

Range 0 to 25 mbar/L<br />

Resolution 0.1 mbar/L<br />

Range 25 to 100 mbar/L<br />

Resolution 0.5 mbar/L<br />

Corresponds to compliance compensation 10000 to 10 mL/mbar<br />

Paediatric mode<br />

Settings for PPS:<br />

Flow Assist (FlowAssist)<br />

Range 0 to 30 mbar/L/s<br />

Resolution 0.5 mbar/(L/s)<br />

Range 30 to 100 mbar/L/s<br />

Resolution 5 mbar/L/s<br />

Corresponds to resistance compensation 0 to 100 mbar/L/s<br />

Volume Assist (Vol.Assist)<br />

Range 0 to 100 mbar/L<br />

Resolution 1 mbar/L<br />

Range 100 to 1000 mbar/L<br />

Resolution 10 mbar/L<br />

Corresponds to compliance compensation 1000 to 1 mL/mbar<br />

O2 Therapy (optional)<br />

Settings<br />

Continuous Flow 2 to 50 L/min<br />

O2 concentration 21 to 100 Vol.%<br />

Accuracy ±5 % of set value, or ±2 Vol.%,<br />

whichever is greater.<br />