PLV*<br />

Pressure Limited Ventilation<br />

Supplementary function for variable limitation<br />

of pressure peaks using the Pmax<br />

pressure limit in ventilation modes IPPV<br />

and SIMV.<br />

The tidal volume remains constant as<br />

long as the pressure curve shows a plateau<br />

and the flow curve shows a brief<br />

flow pause between inspiration and<br />

expiration.<br />

● Activate/deactivate pressure limitation<br />

Pmax, see "Configuration, Set<br />

initial values for O2, I:E...", page 140.<br />

1 When pressure limitation is active,<br />

the value for Pmax is shown in the realtime<br />

curve PAW (t) as a blue line.<br />

2 The screen knob »Pmax« is additionally<br />

displayed in the menu »Ventilator<br />

Settings«.<br />

Volume monitoring is active. The alarm<br />

"Vol. not constant, pressure limited!!" is<br />

automatically triggered if the tidal volume<br />

VT can no longer be applied. This<br />

visual and acoustic alarm can be reset<br />

via the screen key »Alarm Reset« at the<br />

top of the screen until the cause of the<br />

alarm is remedied.<br />

Set PLV with »Pmax«.<br />

* For a detailed description of PLV, please<br />

refer to page 208.<br />

PAW<br />

Flow<br />

Insp.<br />

pause<br />

Tinsp Te<br />

Setting special functions<br />

PLV<br />

Instructions for Use <strong>EvitaXL</strong> 75<br />

PPlat<br />

1<br />

f<br />

Insp. flow<br />

➁<br />

➀<br />

Pmax<br />

(configurable)<br />

t<br />

PEEP<br />

t<br />

005<br />

195