Sensors<br />

Flow sensor calibration<br />

Sensors<br />

<strong>EvitaXL</strong> uses the following sensors for measuring and<br />

monitoring:<br />

— Flow sensor<br />

— Pressure sensors<br />

— O2 sensor<br />

— CO2 sensor (optional)<br />

The last sensor calibration values obtained are saved until the<br />

sensors are calibrated again, even if the ventilator is switched<br />

off in the meantime.<br />

The pressure sensors for measuring the airway pressure are<br />

calibrated automatically.<br />

The flow sensor and O2 sensor are calibrated automatically<br />

once per day.<br />

The flow sensor can be calibrated at any time, even during<br />

ventilation.<br />

The O2 sensor can be calibrated at any time, even during ventilation.<br />

This does not influence the applied O2 concentration.<br />

Calibration of the CO2 sensor (optional) can be checked during<br />

ventilation.<br />

Flow sensor calibration<br />

— After replacing the flow sensor.<br />

The flow sensor is automatically cleaned by <strong>EvitaXL</strong> before it is<br />

calibrated.<br />

It is automatically cleaned and calibrated by <strong>EvitaXL</strong> after use<br />

of the medicament nebuliser.<br />

● Flammable gases (e.g. alcohol vapours after disinfection)<br />

must be avoided.<br />

● Flow sensors which have been disinfected in ethanol<br />

must be allowed to dry in air for at least 30 minutes.<br />

106 Instructions for Use <strong>EvitaXL</strong>