1 Hang the hinged arm from the rail on the left-hand side and<br />

tighten screws.<br />

● Fit the ventilation hoses. Check the hose lengths (metres).<br />

2 Turn ports in direction of hoses.<br />

3 Install water traps in vertical position.<br />

4 Connect the Y-piece, with the rubber sleeve of the Y-piece<br />

on the inspiratory side.<br />

5 Insert the Y-piece in the opening of the hinged arm.<br />

Fitting temperature sensor<br />

6 Push sensor as far as it will go into the rubber sleeve on the<br />

inspiratory side of the Y-piece. Align the Y-piece so that the<br />

sensor is at the top in order to avoid condensation in the<br />

sensor.<br />

7 Attach the sensor cable with hose clips.<br />

8 Insert the connector for the temperature sensor into the<br />

socket »Temp « at the rear of <strong>EvitaXL</strong>.<br />

Preparing for use<br />

Ventilating adults and children<br />

Instructions for Use <strong>EvitaXL</strong> 27<br />

5<br />

4<br />

6<br />

0.6m<br />

0.6m<br />

3<br />

0.6m<br />

0.4m<br />

1<br />

7<br />

D<br />

0.4m<br />

8<br />

2<br />

Evita XL<br />

079<br />

072<br />

080