EvitaXL
EvitaXL
EvitaXL
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Ventilation Modes<br />
Volume-controlled ventilation with PLV and AutoFlow ®<br />
Ventilation Modes<br />
Volume-controlled ventilation with PLV and<br />
AutoFlow ®<br />
AutoFlow is a new additional function that regulates inspiratory<br />
flow during the mandatory ventilation stroke in the constant-volume<br />
ventilation modes IPPV, SIMV and MMV. To explain the<br />
improvement achieved by this function, the conventional methods<br />
are explained first:<br />
Classic constant-volume mandatory ventilation stroke<br />
In mandatory ventilation strokes without AutoFlow, the<br />
»Insp.Flow« parameter restricts the inspiration flow.<br />
If the inspiration flow is so high that the set tidal volume VT is<br />
attained before the inspiration time Tinsp has fully elapsed, the<br />
inspiration valve closes, and the breathing gas supply stops.<br />
The expiration valve also remains closed until the end of the<br />
inspiration time Tinsp. This phase, the inspiratory pause, can be<br />
identified in the curve PAW (t) as the plateau PPlat.<br />
This type of mandatory ventilation stroke, which for technical<br />
reasons is found in the same form in almost all intensive care<br />
ventilators, has two serious drawbacks:<br />
— If the lungs are extremely non-homogeneous, the pressure<br />
peaks can lead to the overdistension of specific lung areas,<br />
and<br />
— the limited inspiration flow and closed inspiration and expiration<br />
valves during the inspiratory pause can cause the<br />
patient to "fight" the machine.<br />
Manual pressure limitation Pmax<br />
<strong>EvitaXL</strong> can prevent pressure peaks, while maintaining the set<br />
tidal volume VT, by means of the pressure limit Pmax. The tidal<br />
volume VT remains constant as long as a pressure plateau<br />
PPlat is still detectable and the flow curve shows a brief zero<br />
flow between inspiration and expiration.<br />
<strong>EvitaXL</strong> performs this function by reducing the Insp. Flow on<br />
reaching the set Pmax value. If the tidal volume VT can no<br />
longer be attained with the selected pressure Pmax, due to<br />
reduced compliance, the alarm "Volume not constant" is automatically<br />
generated.<br />
208 Instructions for Use <strong>EvitaXL</strong><br />
PAW<br />
Flow<br />
Insp.<br />
pause<br />
PPlat<br />
Tinsp Te<br />
1<br />
f<br />
Insp. flow<br />
Pmax<br />
(configurable)<br />
t<br />
PEEP<br />
t<br />
005