Mains power/DC power operation<br />

Operation with external battery (optional)<br />

Operation with external battery (optional)<br />

Connecting an external battery<br />

If the mains power supply fails, <strong>EvitaXL</strong> immediately switches<br />

over to the on-line external battery.<br />

The LED next to the symbol for external battery lights up<br />

yellow.<br />

The green LED next to the symbol for external battery goes<br />

out, as it is no longer fully charged.<br />

The changeover to external battery power is effected without<br />

an alarm message.<br />

The time available for operation with external battery depends<br />

on the battery charge state and the type of battery connected.<br />

If the external battery is exhausted, <strong>EvitaXL</strong> automatically<br />

switches over to the integrated battery and generates an<br />

alarm.<br />

When the mains power is restored, <strong>EvitaXL</strong> automatically<br />

switches back to operation with mains power.<br />

The integrated battery is not recharged while the ventilator<br />

is being powered by an external battery!<br />

For this reason:<br />

● Recharge the integrated and external batteries as soon as<br />

possible, see "Charging the batteries", page 171.<br />

Mains-powered equipment must not be connected to the<br />

DC socket!<br />