CPAP-ASB<br />
Continuous Positive Airway Pressure<br />
Assisted Spontaneous Breathing*<br />
Spontaneous breathing at a raised pressure<br />
level in order to increase the functional<br />
residual capacity (FRC).<br />
Spontaneous breathing can be assisted<br />
with additional pressure by ASB.<br />
Set the pattern of ventilation for CPAP<br />
and CPAP/ASB with the following ventilation<br />
parameters:<br />
O2 concentration »O2«<br />
Positive end-expiratory pressure »PEEP«<br />
Pressure support »PASB«<br />
Pressure rise time »Ramp«<br />
--------------------------------------------------------------------------------<br />
To set:<br />
● Touch the appropriate screen knob.<br />
● Adjust to the desired value<br />
= turn rotary knob.<br />
● Confirm setting = press rotary knob.<br />
--------------------------------------------------------------------------------<br />
To view additional text information on<br />
CPAP, CPAP/ASB:<br />
● Touch the screen key »? q «.<br />
CPAP, CPAP/ASB can be supplemented<br />
to include the following special<br />
functions:<br />
— Apnoea ventilation, page 69.<br />
— ATC, page 72.<br />
— Flowtrigger, page 68.<br />
These functions can be enabled in »Add.<br />
settings«.<br />
Setting alarm limits, see page 79.<br />
* For a detailed description of CPAP/ASB,<br />
please refer to page 216.<br />
PAW<br />
Flow<br />
fast<br />
rise time<br />
Spontaneous breathing<br />
with ASB pressure support<br />
slow<br />
rise time<br />
Setting Ventilation Modes<br />
CPAP-ASB<br />
Instructions for Use <strong>EvitaXL</strong> 63<br />
PEEP<br />
t<br />
t<br />
010<br />
115