CPAP-ASB<br />

Continuous Positive Airway Pressure<br />

Assisted Spontaneous Breathing*<br />

Spontaneous breathing at a raised pressure<br />

level in order to increase the functional<br />

residual capacity (FRC).<br />

Spontaneous breathing can be assisted<br />

with additional pressure by ASB.<br />

Set the pattern of ventilation for CPAP<br />

and CPAP/ASB with the following ventilation<br />

parameters:<br />

O2 concentration »O2«<br />

Positive end-expiratory pressure »PEEP«<br />

Pressure support »PASB«<br />

Pressure rise time »Ramp«<br />

--------------------------------------------------------------------------------<br />

To set:<br />

● Touch the appropriate screen knob.<br />

● Adjust to the desired value<br />

= turn rotary knob.<br />

● Confirm setting = press rotary knob.<br />

--------------------------------------------------------------------------------<br />

To view additional text information on<br />

CPAP, CPAP/ASB:<br />

● Touch the screen key »? q «.<br />

CPAP, CPAP/ASB can be supplemented<br />

to include the following special<br />

functions:<br />

— Apnoea ventilation, page 69.<br />

— ATC, page 72.<br />

— Flowtrigger, page 68.<br />

These functions can be enabled in »Add.<br />

settings«.<br />

Setting alarm limits, see page 79.<br />

* For a detailed description of CPAP/ASB,<br />

please refer to page 216.<br />

PAW<br />

Flow<br />

fast<br />

rise time<br />

Spontaneous breathing<br />

with ASB pressure support<br />

slow<br />

rise time<br />

Setting Ventilation Modes<br />

CPAP-ASB<br />

Instructions for Use <strong>EvitaXL</strong> 63<br />

PEEP<br />

t<br />

t<br />

010<br />

115