Preparing for use<br />

Fitting O2 sensor capsule<br />

Fitting O2 sensor capsule<br />

— when using the system for the first time<br />

— when the display reads:<br />

O2 measurement inop<br />

— when calibration can no longer be performed<br />

● Ensure device is in standby or ventilator is switched off.<br />

● Tilt the control unit upwards, pressing the segments on the<br />

right and left at the same time.<br />

1 Turn the inspiratory port to the left.<br />

2 Use coin to loosen screw, and remove protective cover.<br />

3 Loosen the two knurled screws and open the sensor<br />

housing.<br />

4 Remove old sensor capsule and fit a new capsule. The end<br />

with the circular tracks remains visible.<br />

3 Close the sensor housing securely with the two knurled<br />

screws.<br />

2 Screw protective cover back in place.<br />

● Dispose of the used O2 sensor capsule, page 168.<br />

24 Instructions for Use <strong>EvitaXL</strong><br />

1<br />

4<br />

3<br />

2<br />

071