Display logbook<br />
Changes, events and alarms are registered<br />
by <strong>EvitaXL</strong> and listed in chronological<br />
order with the date and time of<br />
occurrence.<br />
Changes are displayed with the former<br />
and new settings (example: 5 mbar – ><br />
7 mbar).<br />
Events include, for instance, use of a<br />
medicament nebuliser, Flow calibration,<br />
etc.<br />
Alarms are registered in the form displayed<br />
by <strong>EvitaXL</strong> at the time of occurrence.<br />
Other alarms which are triggered<br />
with the displayed alarm but are not<br />
themselves displayed in the field for<br />
alarm messages are identified by an<br />
asterisk (*) preceding the entry in the<br />
logbook.<br />
To display the logbook:<br />
● Touch the function key »Data...«,<br />
the »Data« menu is displayed.<br />
● Touch the screen key »Logbook« and<br />
<strong>EvitaXL</strong> displays the logbook.<br />
When a time is highlighted in the trend<br />
display (page 90), the line corresponding<br />
to that time is also marked in the logbook.<br />
For the highlighted line, <strong>EvitaXL</strong> displays<br />
the complete list of new settings for<br />
every registered change in parameters<br />
for the ventilation mode effective at that<br />
time.<br />
To view all the settings for another line:<br />
● Turn the rotary knob to select the<br />
required line.<br />
● Touch the screen key »x« to close the<br />
logbook.<br />
Display measured values<br />
Display logbook<br />
Instructions for Use <strong>EvitaXL</strong> 89<br />