Setting alarm limits<br />

Suppress alarm tone<br />

Power failure alarm<br />

Suppress alarm tone<br />

for max. 2 minutes:<br />

1 Press key »g Alarm Silence«, its yellow<br />

LED lights up.<br />

The acoustic alarm is suppressed for<br />

two minutes.<br />

2 The remaining time is displayed on<br />

the screen.<br />

If the fault triggering the alarm has not<br />

been remedied after 2 minutes, the<br />

acoustic alarm sounds again.<br />

If the acoustic alarm is to be re-activated<br />

prematurely:<br />

1 Press key »g Alarm Silence« again,<br />

its LED goes out.<br />

Acknowledge:<br />

See "Fault – Cause – Remedy",<br />

page 143, for alarms which can be reset<br />

via the Alarm Reset key.<br />

● Touch the screen key »Alarm Reset«,<br />

confirm = press rotary knob.<br />

Power failure alarm<br />

If the loudspeaker for acoustic alarm signalling<br />

fails on account of a defect, a<br />

continuous tone will be generated by an<br />

auxiliary alarm. This continuous tone also<br />

draws attention to a power failure (see<br />

page 30, in the event of an interruption<br />

in the power supply).<br />

➁<br />

D<br />

82 Instructions for Use <strong>EvitaXL</strong><br />

<strong>EvitaXL</strong><br />

1<br />

032<br />

214