Mains power/DC power operation<br />
Charging times<br />
Battery maintenance<br />
Charging times<br />
The specified charging times apply when batteries are<br />
recharged immediately after a full discharge.<br />
The charging time may be significantly longer if the batteries<br />
have discharged several times in succession without being<br />
fully recharged on mains power in the meantime.<br />
The batteries must be fully operational.<br />
Charge indication and battery charge state<br />
Charging of the integrated and external batteries is interrupted<br />
when the charging current drops to a very low value upon<br />
reaching the end of charge. The battery is considered to be<br />
fully charged and this is indicated by a green battery symbol.<br />
The battery capacity actually available at the end of the charging<br />
process depends, among other things, on the condition of<br />
the battery and on the ambient temperature. The battery<br />
capacity and condition of the battery cannot be determined by<br />
the DC power supply.<br />
The green battery symbol indicates that the battery is fully<br />
charged. Even though the green symbol lights up, the capacity<br />
of old or defective batteries may be so small as to permit operation<br />
of <strong>EvitaXL</strong> for no more than a few minutes.<br />
Sufficient battery capacity is always required.<br />
See "Battery maintenance"!<br />
Battery maintenance<br />
To ensure maximum battery life:<br />
● The battery should always be fully charged and where possible<br />
never left uncharged.<br />
If the DC power supply is not used:<br />
● Connect <strong>EvitaXL</strong> to mains power after not more than one<br />
month and switch it on for at least two hours in order to<br />
recharge the integrated battery.<br />
● Then fully recharge any external batteries which may be<br />
connected.<br />
If the battery is left uncharged for more than one month:<br />
● The integrated and external batteries should be disconnected<br />
from <strong>EvitaXL</strong> by experts in order to reduce spontaneous<br />
discharge by the batteries.<br />
Before reconnecting the integrated and external batteries,<br />
check that their capacity is still adequate. They may have<br />
become exhausted or damaged as a result of prolonged<br />
storage.<br />
Batteries are wear parts. They must be replaced in accordance<br />
with the extent to which they are used.<br />
● Exhaustive discharges should be avoided, as they lead<br />
to premature wear!<br />
Batteries are wear parts. Their capacity must be checked<br />
regularly and the batteries replaced if necessary.<br />