Mains power/DC power operation<br />

Charging times<br />

Battery maintenance<br />

Charging times<br />

The specified charging times apply when batteries are<br />

recharged immediately after a full discharge.<br />

The charging time may be significantly longer if the batteries<br />

have discharged several times in succession without being<br />

fully recharged on mains power in the meantime.<br />

The batteries must be fully operational.<br />

Charge indication and battery charge state<br />

Charging of the integrated and external batteries is interrupted<br />

when the charging current drops to a very low value upon<br />

reaching the end of charge. The battery is considered to be<br />

fully charged and this is indicated by a green battery symbol.<br />

The battery capacity actually available at the end of the charging<br />

process depends, among other things, on the condition of<br />

the battery and on the ambient temperature. The battery<br />

capacity and condition of the battery cannot be determined by<br />

the DC power supply.<br />

The green battery symbol indicates that the battery is fully<br />

charged. Even though the green symbol lights up, the capacity<br />

of old or defective batteries may be so small as to permit operation<br />

of <strong>EvitaXL</strong> for no more than a few minutes.<br />

Sufficient battery capacity is always required.<br />

See "Battery maintenance"!<br />

Battery maintenance<br />

To ensure maximum battery life:<br />

● The battery should always be fully charged and where possible<br />

never left uncharged.<br />

If the DC power supply is not used:<br />

● Connect <strong>EvitaXL</strong> to mains power after not more than one<br />

month and switch it on for at least two hours in order to<br />

recharge the integrated battery.<br />

● Then fully recharge any external batteries which may be<br />

connected.<br />

If the battery is left uncharged for more than one month:<br />

● The integrated and external batteries should be disconnected<br />

from <strong>EvitaXL</strong> by experts in order to reduce spontaneous<br />

discharge by the batteries.<br />

Before reconnecting the integrated and external batteries,<br />

check that their capacity is still adequate. They may have<br />

become exhausted or damaged as a result of prolonged<br />

storage.<br />

Batteries are wear parts. They must be replaced in accordance<br />

with the extent to which they are used.<br />

● Exhaustive discharges should be avoided, as they lead<br />

to premature wear!<br />

Batteries are wear parts. Their capacity must be checked<br />

regularly and the batteries replaced if necessary.<br />