Fault – Cause – Remedy<br />
Message Cause Remedy<br />
Flow Sensor? !!! Flow sensor not fully inserted in rubber lip<br />
of expiration valve.<br />
Hose kinked !! The pressure at the inspiratory port is<br />
greater than 30 mbar, e.g. due to a kinked<br />
or blocked hose, or a blocked patient<br />
mask.<br />
High frequency !!! Patient is breathing at a high spontaneous<br />
frequency.<br />
ILV sync. inop.<br />
Message on both devices<br />
Insert flow sensor correctly.<br />
Check patient hoses, check patient mask.<br />
Check condition of patient.<br />
Check pattern of ventilation or<br />
spontaneous breathing frequency.<br />
Check hose system for water (auto<br />
triggering).<br />
Correct alarm limit if neccessary.<br />
!!! Frequency on master device less than<br />
4 breaths per minute.<br />
Set a higher frequency.<br />
Fault in communication between the units. The ventilator functions are not affected,<br />
but there is no communication between<br />
the two units. Replace ventilator to<br />
continue use of ILV.<br />
Call DrägerService.<br />
Insp. hold interrupted ! The »Insp. hold« key was operated for<br />
longer than 15 seconds.<br />
Int. battery discharged !!! The ventilator is being powered by DC<br />
integrated battery due to the absence of<br />
mains power and an external battery. The<br />
time for operation with power from the<br />
integrated battery has expired.<br />
Int. battery in operation ! The ventilator is being powered by DC<br />
integrated battery due to the absence of<br />
mains power and an external battery.<br />
The maximum remaining time from the<br />
integrated battery is 10 minutes.<br />
Int. battery only 2 minutes<br />
left<br />
!! The ventilator is being powered by DC<br />
integrated battery due to the absence of<br />
mains power and an external battery. The<br />
maximum remaining time from the<br />
integrated battery is 2 minutes.<br />
Key O2 Y suction overused? !! Key has been pressed frequently in a short<br />
period of time.<br />
Key overused? !! Due to frequent key use, the screen<br />
contents of the display are repeatedly<br />
redrawn.<br />
Brief communication failure between<br />
display processor and main processor.<br />
Release »Insp. hold« key.<br />
Connect ventilator immediately to mains<br />
power supply or to a fully charged external<br />
battery.<br />
Connect ventilator to the mains power<br />
supply or to a fully charged battery within<br />
10 minutes.<br />
Connect ventilator to the mains power<br />
supply or to a fully charged battery within<br />
2 minutes.<br />
Acknowledge message by pressing<br />
»Alarm Reset« key and confirm. If this<br />
message appears repeatedly:<br />
Disconnect patient from the device and<br />
continue ventilation immediately with<br />
another independent ventilator.<br />
Call DrägerService.<br />
Acknowledge message by pressing<br />
»Alarm Reset« key and confirm.<br />
Acknowledge message by pressing<br />
»Alarm Reset« key and confirm. If this<br />
message appears repeatedly:<br />
Disconnect patient from the device and<br />
continue ventilation immediately with<br />
another independent ventilator.<br />
Call DrägerService<br />
148 Instructions for Use <strong>EvitaXL</strong>