Monitoring<br />
Expiratory minute volume MV<br />
Upper alarm limit alarm if the upper alarm limit has been exceeded<br />
Setting range 41 to 0.1 L/min, in 0.1 L/min steps<br />
Lower alarm limit alarm if the value has fallen below the lower alarm limit<br />
Setting range 0.01 to 40 L/min, in 0.1 L/min steps<br />
Technical Data<br />
Monitoring<br />
Airway pressure PAW<br />
Upper alarm limit alarm if the "PAW high" value is exceeded<br />
Setting range 10 to 100 mbar<br />
Lower alarm limit alarm if the value "PEEP +5 mbar" (coupled with the PEEP set value) is not<br />
exceeded for at least 96 ms in 2 successive ventilation strokes.<br />
Insp. O2 concentration FiO2<br />
Upper alarm limit alarm if the upper alarm limit is exceeded for at least 20 seconds<br />
Lower alarm limit alarm if the value falls below the lower alarm limit for at least 20 seconds<br />
Range both alarm limits are automatically allocated to the set value:<br />
under 60 Vol.% at ±4 Vol.%<br />
over 60 Vol.% at ±6 Vol.%<br />
End-expiratory CO2 concentration<br />
etCO2 (optional)<br />
Upper alarm limit alarm if the upper alarm limit has been exceeded<br />
Setting range 0 to 100 mmHg<br />
or<br />
0 to 15 kPa<br />
Lower alarm limit alarm if the value has fallen below the lower alarm limit<br />
Setting range 0 to 99 mmHg<br />
or<br />
0 to 14 kPa<br />
Insp. breathing gas temperature<br />
Upper alarm limit alarm on reaching 40 o C<br />
(<strong>EvitaXL</strong> can also be used without a temperature sensor, if it is not connected<br />
when <strong>EvitaXL</strong> is switched on)<br />
Tachypnoea monitoring fspn<br />
Alarm during spontaneous breathing, when the spontaneous breathing frequency<br />
has been exceeded<br />
Setting range 5 to 120/min<br />
Volume monitoring<br />
Lower alarm limit alarm if the set tidal volume VT (coupled with the set value VT) has not been<br />
supplied<br />
Upper alarm limit alarm if the applied tidal volume exceeds the value of the alarm limit, inspiration<br />
is interrupted and the expiration valve is opened<br />
Setting range 21 to 4000 mL<br />
Apnoea alarm time<br />
Alarm if no breathing activity is detected<br />
Setting range 5 to 60 seconds, can be set in 1 second increments<br />
Instructions for Use <strong>EvitaXL</strong> 195