The difference between measured values under NTPD and<br />

BTPS conditions is typically approx. 12 %.<br />

Example: heating up 500 mL tidal volume NTPD to 37 o C and<br />

moistening it to 100 % rel. humidity produces 564 mL BTPS.<br />

<strong>EvitaXL</strong> doses the tidal volume so that the set tidal volume<br />

under BTPS is effective in the lung.<br />

The expiratory measurement is performed on the basis of saturated<br />

gases at 30 o C.<br />

Measurement principles<br />

Flow Measurement<br />

The expiratory flow is measured with a hot-wire anemometer.<br />

The energy needed to keep the hot wire at a temperature of<br />

180 oC represents the flow flowing through the sensor and<br />

cooling the hot wire.<br />

O2 measurement<br />

The oxygen measurement is based on the principle of a galvanic<br />

cell. The monitored gas diffuses through a membrane<br />

into the electrolyte in the sensor. The electrolyte contains a<br />

working electrode and a reference electrode. The oxygen is<br />

reduced electrochemically and the resultant current is proportional<br />

to the O2 partial pressure in the gas.<br />

CO2 measurement<br />

CO2 is measured in the main stream and is based on an<br />

absorption measurement. A spectrum is output by a light<br />

source while two detectors register the characteristic absorption<br />

spectrum and generate electrical signals as a function of<br />

the CO2 concentration. These signals are evaluated and displayed.<br />

Condensation is prevented by heating the CO2<br />

measuring unit.<br />

Measurements<br />

Flow Measurement<br />

Measurement principles<br />

Instructions for Use <strong>EvitaXL</strong> 219