Connecting the nurse call (optional)<br />

Socket on the rear of <strong>EvitaXL</strong> for connecting alarm signals to a<br />

central alarm station in the hospital.<br />

● The kit may only be installed by specialists.<br />

For details of the characteristics, refer to the technical data,<br />

page 197.<br />

● The 6-pin round DIN plug (female connector) must be connected<br />

to the lead for the central alarm station in the hospital<br />

by a specialist.<br />

Connection 3 to 5 makes and the nurse call is activated as<br />

soon as <strong>EvitaXL</strong> signals an alarm.<br />

The central hospital alarm system may only be connected<br />

to the nurse call if <strong>EvitaXL</strong> is connected to the mains<br />

power supply via a mains power cable or if it has been<br />

earthed via the earth connection on the back of <strong>EvitaXL</strong>.<br />

Electric power may pose a hazard in all other cases.<br />

1 Plug the connector into the » « socket on the rear and<br />

screw into place.<br />

Only alarm messages of the highest priority (see<br />

page 80) are transmitted via nurse call.<br />

Alarm messages are displayed in the top line of the screen in<br />

red and with three exclamation marks, see page 80.<br />

Caution and advisory messages are not transmitted.<br />

The nurse call is also activated when the internal acoustic generator<br />

in the ventilator is defective.<br />

● Check correct operation of connected nurse call system.<br />

Connection of a nurse call does not relieve staff of their<br />

duty to check the monitoring on the <strong>EvitaXL</strong> screen at regular<br />

intervals.<br />

● Screen displays must be checked regularly.<br />

A fault in any of the components in the link between<br />

nurse call and central hospital alarm system (e.g. in the<br />

electronics for nurse call in <strong>EvitaXL</strong>, in the <strong>EvitaXL</strong> power<br />

supply or in the alarm generator of the central hospital<br />

alarm system, etc.) may result in failure of the nurse call.<br />

The hospital connections to the central alarm typically use only<br />

one channel. The electronics for nurse call consequently also<br />

uses only one channel.<br />

Preparing for use<br />

Connecting the nurse call (optional)<br />

1 5<br />

3<br />

Instructions for Use <strong>EvitaXL</strong> 33<br />

1<br />

087