Resetting CO2 calibration<br />

— If calibration was not successful or if<br />

there were problems during calibration,<br />

the sensor can be reset to the<br />

factory reset values.<br />

In the menu »CO2«:<br />

● Touch the screen key »Calibration«.<br />

● Touch the screen key »Reset Cal.«.<br />

The calibration is reset by <strong>EvitaXL</strong> after<br />

approx. 5 seconds and the factory-set<br />

calibration value is now effective again.<br />

● Recover the correct calibration as<br />

soon as possible!<br />

Sensors<br />

Zero/check/calibrate CO2 sensor<br />

Instructions for Use <strong>EvitaXL</strong> 117<br />