Clean and disinfect equipment and/or components before<br />
any maintenance procedures – and before returning for<br />
repair!<br />
O2 sensor Replace sensor capsule in event of display<br />
message:<br />
O2 measurement inop<br />
and if calibration is no longer possible.<br />
Disposal, see page 168.<br />
Ambient air filter and<br />
Cooling air filter<br />
Clean or replace after 4 weeks, see page 167.<br />
Replace after 1 year.<br />
Dispose of with normal domestic waste.<br />
Filters in the compressed gas inlets To be replaced by trained service personnel<br />
every 2 years.<br />
Covering grid for heating patient part Clean after 4 weeks.<br />
Dirt blocks the air inlet and reduces the heating<br />
output.<br />
Lithium battery for data protection To be replaced by trained service personnel<br />
every 2 years.<br />
Disposal, see page 168.<br />
Integrated battery of DC power<br />
supply<br />
Must be serviced in conjunction with halfyearly<br />
inspections. To be replaced by trained<br />
service personnel every two years at the latest.<br />
Check capacity every six months!<br />
Replace battery if necessary.<br />
External battery (optional) Must be serviced in conjunction with halfyearly<br />
inspections.<br />
Check capacity every six months!<br />
Replace battery if necessary.<br />
Clock module To be replaced by trained service personnel<br />
after 6 years.<br />
Pressure reducer To be replaced every 6 years by<br />
DrägerService.<br />
Equipment inspection and service Every 6 months by trained service personnel.<br />
Safety checks (only applies to the<br />
Federal Republic of Germany)<br />
Every 6 months in accordance with Art. 6<br />
MPBetreibV (see "Safety checks" sheet)<br />
166 Instructions for Use <strong>EvitaXL</strong>