Sensors<br />

External flow compensation<br />

In the »Flow« menu:<br />

● Touch screen key »measure«, the<br />

key turns yellow. <strong>EvitaXL</strong> measures<br />

the external flow and displays it in the<br />

menu with the date and time.<br />

The following message is displayed<br />

while measurement is in progress:<br />

Measuring external flow<br />

Calculation of the external flow is interrupted<br />

by <strong>EvitaXL</strong> if it exceeds 12 L/min<br />

or if the flow measurement function is<br />

defective.<br />

When the external flow has been calculated<br />

successfully, it is taken into<br />

account automatically and the screen<br />

key »On« appears in green.<br />

If an external flow is not applied:<br />

● Touch the screen key »Off«, the key<br />

turns yellow.<br />

● Confirm = press rotary knob, the key<br />

turns pale green.<br />

Once the external flow has been calculated<br />

by <strong>EvitaXL</strong>, it can be taken into<br />

account at any time:<br />

● Touch the screen key »On« in the<br />

»Flow« menu, the key turns yellow.<br />

● Confirm = press the rotary knob, the<br />

key turns green.<br />

If the external flow changes:<br />

● Touch the screen key »measure«<br />

again and the new external flow is calculated<br />

by <strong>EvitaXL</strong>.<br />

108 Instructions for Use <strong>EvitaXL</strong><br />

138