Technical Data<br />

Computed value displays<br />

Computed value displays<br />

Compliance C<br />

Range 0 to 300 mL/mbar<br />

Resolution<br />

Range 0 to 99.9 mL/mbar 0.1 mL/mbar<br />

Range 100 to 300 mL/mbar 1 mL/mbar<br />

Accuracy ±20 % of measured value*<br />

Resistance R<br />

Range 0 to 600 mbar/L/s<br />

Resolution<br />

Range 0 to 99.9 mbar/L/s 0.1 mbar/L/s<br />

Range 100 to 600 mbar/L/s 1 mbar/L/s<br />

Accuracy ±20 % of measured value**<br />

Leakage minute volume MVleak<br />

Range 0 to 99 L/min, BTPS<br />

Resolution 0.1 L/min or for values less than 0.1 L/min: 0.01 L/min<br />

Accuracy ±18 % of measured value<br />

T10...90 approx. 35 s<br />

Rapid Shallow Breathing RSB<br />

Range 0 to 9999 1/(min x L)<br />

Resolution 1/(min x L)<br />

Accuracy see measurement of VT and f<br />

Negative Inspiratory Force NIF<br />

Range –45 to 0 mbar<br />

Resolution 1 mbar<br />

Accuracy ±2 mbar<br />

Curve displays:<br />

Airway pressure PAW (t) –10 to 100 mbar<br />

Flow (t) –200 to 200 L/min<br />

Volume V (t) 0 to 2000 mL<br />

Exp. CO2 concentration FCO2 0 to 100 mmHg or<br />

0 to 14 kPa or<br />

0 to 15 Vol.%<br />

Occlusion pressure P 0.1 0 to 25 mbar<br />

* C-values may be considerably falsified as spontaneous breathing increases;<br />

compliance with the measuring accuracy therefore cannot be guaranteed for spontaneous breathing.<br />

** R-values may be considerably falsified as spontaneous breathing increases;<br />

compliance with the measuring accuracy therefore cannot be guaranteed for spontaneous breathing.<br />

194 Instructions for Use <strong>EvitaXL</strong>